Automated Pipetting, Diluting And Specimen Processing Workstations For Flow Cytometric Analysis
Clinical Chemistry · 21 CFR 862.2750
Flow cytometer specimen processors are automated or semi-automated workstations designed to be used alone or in combination with manual methods to pipette, dilute and process human specimens in preparation for flow cytometric analysis.
Recent MDR events
181 adverse event reports on file
- BD FACS¿ SAMPLE PREP ASSISTANT IIIFeb 6, 2026MalfunctionBecton, Dickinson And Company, Bd Biosciences3027394506-2026-00002
- BD FACS¿ SAMPLE PREP ASSISTANT IIIDec 12, 2025InjuryBecton, Dickinson And Company, Bd Biosciences3027394506-2025-00048
- BD FACS¿ SAMPLE PREP ASSISTANT IIISep 25, 2025MalfunctionBecton, Dickinson And Company, Bd Biosciences3027394506-2025-00041
- BD FACS¿ SAMPLE PREP ASSISTANT IIIMay 19, 2025MalfunctionBecton, Dickinson And Company, Bd Biosciences3027394506-2025-00023
- SYSMEX PS-10Feb 6, 2025MalfunctionSysmex Corporation1000515253-2025-00004
- SYSMEX PS-10Jan 10, 2025MalfunctionSysmex Corporation1000515253-2025-00002
- SYSMEX PS-10Jan 9, 2025MalfunctionSysmex Corporation1000515253-2025-00001
- BD FACSDUET¿ SAMPLE PREPAug 21, 2024MalfunctionBecton Dickinson And Company Bd Biosciences2916837-2024-00045
- BD FACS¿ SAMPLE PREP ASSISTANT IIIMar 22, 2024MalfunctionBecton, Dickinson And Company, Bd Biosciences2916837-2024-00024
- BD FACS¿ SAMPLE PREP ASSISTANT IIIJan 18, 2024MalfunctionBecton, Dickinson And Company, Bd Biosciences2916837-2024-00001
Event type breakdown
Recalls
10 recall events for this product code
- Beckman Coulter, Inc.Open, ClassifiedFeb 2, 2026
CellMek SPS Sample Preparation System, REF: C44603, with software
Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.
Z-1235-2026 · View FDA record ↗ - Beckman Coulter, Inc.Open, ClassifiedOct 30, 2025
CellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU) revisions AD and prior. These CellMek
Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment.
Z-0412-2026 · View FDA record ↗ - Sysmex America, Inc.Open, ClassifiedFeb 5, 2025
Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341.
Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.
Z-1113-2025 · View FDA record ↗ - Becton, Dickinson and Company, BD Bio SciencesOpen, ClassifiedOct 3, 2024
BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSDuet Sample Preparation System (Base), REF: 663128, 662588, used with software V
Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated using Trucount Tubes: increased volume in tube1 may result in 20% higher absolute counts in tube 1 than tube 2; could result in additional blood collection, delayed results/patient therapy.
Z-0009-2025 · View FDA record ↗ - Becton, Dickinson and Company, BD BiosciencesTerminatedUnknown
BD FACS Sample Prep Assistant (SPA) II - Product Usage: K050191: The BD FACSCanto system with BD FACSCanto clinical software when used with the BD FACS Sample P
The recalling firm confirmed that BD FACS Sample Prep Assistant (SPA) II and III users may encounter probes with unacceptable wear prior to reaching the 2500 piercings guideline indicated in the Instructions for Use (IFU). When the probe is worn, particles from the rubber cap are more likely to clog the probe, resulting in blocked or inadequate sample dispense.
Z-2356-2020 · View FDA record ↗ - Becton, Dickinson and Company, BD BiosciencesTerminatedUnknown
BD FACS Sample Prep Assistant (SPA) II Upgrade - Product Usage: K050191: The BD FACSCanto system with BD FACSCanto clinical software when used with the BD FACS
The recalling firm confirmed that SPA II and III users may encounter probes with unacceptable wear prior to reaching the 2500 piercings guideline indicated in the Instructions for Use (IFU). When the probe is worn, particles from the rubber cap are more likely to clog the probe, resulting in blocked or inadequate sample dispense.
Z-2357-2020 · View FDA record ↗ - Becton, Dickinson and Company, BD BiosciencesTerminatedUnknown
BD FACS Sample Prep Assistant (SPA) III - Product Usage: Prep Assistant III is intended to prepare human whole blood for flow cytometric analysis on BD FACSCant
The recalling firm confirmed that SPA II and III users may encounter probes with unacceptable wear prior to reaching the 2500 piercings guideline indicated in the Instructions for Use (IFU). When the probe is worn, particles from the rubber cap are more likely to clog the probe, resulting in blocked or inadequate sample dispense.
Z-2358-2020 · View FDA record ↗ - Sysmex America, Inc.TerminatedUnknown
Sysmex PS-10 Sample Preparation System
Insufficient amount of antibody without an error message or alarm
Z-2687-2020 · View FDA record ↗ - Beckman Coulter Inc.TerminatedUnknown
Coulter PrepPlus, Part Number 286600,
When samples are prepared using the PrepPlus or PrepPlus 2 sample preparation systems and then run on a Flow cytometer, there is an imprecision in the absolute counts on only the first tube in a run.
Z-2689-2017 · View FDA record ↗ - Beckman Coulter Inc.TerminatedUnknown
Coulter PrepPlus 2, Part Number 378600,
When samples are prepared using the PrepPlus or PrepPlus 2 sample preparation systems and then run on a Flow cytometer, there is an imprecision in the absolute counts on only the first tube in a run.
Z-2690-2017 · View FDA record ↗
Registered firms
19 firms currently registered for this product code
Sysmex Corporation
KOBE Hyogo, JP
Becton, Dickinson And Company, Bd Biosciences
Milpitas, CA, US
Sts Corporation
Duluth, GA, US
Sysmex Corporation Techno Park
KOBE Hyogo, JP
Sysmex America, Inc. - Vernon Hills
Vernon Hills, IL, US
Stratec Se
BIRKENFELD, GRAEFENHAUSEN Baden-Wurttemberg, DE
Sysmex America, Inc.
Lincolnshire, IL, US
Tecan Cdmo Solutions Mh, Corp.
MORGAN HILL, CA, US
Labotek Group Corp
Bayamon, PR, US
Stratec Switzerland Ag
Beringen Schaffhausen, CH
Beckman Coulter, Inc.
Miami, FL, US
One Lambda, Inc.
West Hills, CA, US
One Lambda, Inc
Hennigsdorf Brandenburg, DE
Bloodscan Biotech Inc.
Nangang Dist. Taipei, TW
510(k) clearances
3 clearances for this product code