BD FACS¿ SAMPLE PREP ASSISTANT III
Received Feb 6, 2026 · Event occurred Jan 28, 2026
Report 3027394506-2026-00002 · MDR key 24277294
Device
Generic name
Automated Pipetting, Diluting And Specimen Processing Workstations For Flow Cyto
Manufacturer
Becton, Dickinson And Company, Bd BiosciencesCatalog number
647205
Product problems
- Contamination
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE FOLLOWING FIELD WAS UPDATED WITH CORRECTED INFORMATION: H.6 INVESTIGATION SUMMARY: BASED ON THE INVESTIGATION RESULTS, THE REPORTED ISSUE OF THE CUSTOMER EXPERIENCING CONTAMINATION AFTER TESTING WAS CONFIRMED. INVESTIGATION RESULTS THAT WERE PERFORMED ON THE INDICATED FAILURE MODE WERE THE FOLLOWING: THE COMPLAINT TREND AND SERVICE NOTES WERE REVIEWED. THE POTENTIAL CAUSE FOR THE CUSTOMER EXPERIENCING CONTAMINATION WAS LINKED TO THE 4-WAY VALVE. THE ISSUE WAS RESOLVED AFTER THE PART WAS REPLACED. THE REPLACED PART WAS NOT RETURNABLE; THEREFORE, IT WAS DISCARDED. NO FURTHER PROBLEMS WERE NOTED, AND THE INSTRUMENT WAS CONFIRMED TO BE FUNCTIONING AS EXPECTED. ALTHOUGH THE INSTRUMENT WAS USED FOR DIAGNOSTIC TESTING, THE ISSUE WAS RESOLVED BEFORE THE PATIENT SAMPLE RESULTS WERE USED FOR ANY DIAGNOSIS OR TREATMENT.
Description of Event or Problem
IT WAS REPORTED THAT DURING USE OF BD FACS¿ SAMPLE PREP ASSISTANT III, FOREIGN MATTER/CONTAMINATION IN INSTRUMENT THAT IMPACTED PATIENT RESULTS WAS OBSERVED. CONTAMINANT IS IN A POSITION WHERE IT IMPACTED A READOUT TO BE USED CLINICALLY. THERE WAS NO PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED BY THE CUSTOMER THAT SAMPLES NOT BEING PROCESSED CORRECT. CUSTOMER REPLACED SYRINGE BUT ISSUE PERSISTS. PATIENT SAMPLES CHECKLIST: 1. DID THIS CAUSE ERRONEOUS RESULTS ON PATIENT SAMPLES FOR DIAGNOSTIC TESTING? IF YES, GO TO QUESTION #2. IF NO, NO FURTHER QUESTIONS REQUIRED. YES 2. WAS AN ERRONEOUS LABORATORY RESULT FROM THE BDB PRODUCT REPORTED TO A CLINICIAN? IF YES, GO TO QUESTION #3. IF NO, NO FURTHER QUESTIONS REQUIRED. NO CONTAMINATION: 1. WERE PATIENT SAMPLES CONTAMINATED OR WAS THERE CARRYOVER (CROSS-CONTAMINATION/MIXING) BETWEEN PATIENT SAMPLES? IF NO, NO FURTHER QUESTIONS REQUIRED. PATIENT SAMPLES CONTAMINATED, CARRYOVER BETWEEN PATIENT SAMPLES, UNKNOWN - GO TO QUESTION #2. PATIENT SAMPLES CONTAMINATED 2. PLEASE DESCRIBE ANY ADDITIONAL DETAILS: GO TO PATIENT SAMPLES CHECKLIST. SAMPLES WERE NOT PROPERLY PROCESSED AND THERE WAS A RED TINGE (BLOOD) TO THE COMPLETED SAMPLES. CUSTOMER IDENTIFIED THIS IS NOT CORRECT SO THEY STOPPED USING SYSTEM. CUSTOMER THEN MANUALLY PROCESSED SAMPLES BY HAND.
Description of Event or Problem
IT WAS REPORTED THAT DURING USE OF BD FACS¿ SAMPLE PREP ASSISTANT III, FOREIGN MATTER/CONTAMINATION IN INSTRUMENT THAT IMPACTED PATIENT RESULTS WAS OBSERVED. CONTAMINANT IS IN A POSITION WHERE IT IMPACTED A READOUT TO BE USED CLINICALLY. THERE WAS NO PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED BY THE CUSTOMER THAT SAMPLES NOT BEING PROCESSED CORRECT. CUSTOMER REPLACED SYRINGE BUT ISSUE PERSISTS. PATIENT SAMPLES CHECKLIST: 1. DID THIS CAUSE ERRONEOUS RESULTS ON PATIENT SAMPLES FOR DIAGNOSTIC TESTING? IF YES, GO TO QUESTION #2. IF NO, NO FURTHER QUESTIONS REQUIRED. YES. 2. WAS AN ERRONEOUS LABORATORY RESULT FROM THE BDB PRODUCT REPORTED TO A CLINICIAN? IF YES, GO TO QUESTION #3. IF NO, NO FURTHER QUESTIONS REQUIRED. NO. CONTAMINATION: 1. WERE PATIENT SAMPLES CONTAMINATED OR WAS THERE CARRYOVER (CROSS-CONTAMINATION/MIXING) BETWEEN PATIENT SAMPLES? IF NO, NO FURTHER QUESTIONS REQUIRED. PATIENT SAMPLES CONTAMINATED, CARRYOVER BETWEEN PATIENT SAMPLES, UNKNOWN - GO TO QUESTION #2. PATIENT SAMPLES CONTAMINATED 2. PLEASE DESCRIBE ANY ADDITIONAL DETAILS: GO TO PATIENT SAMPLES CHECKLIST. SAMPLES WERE NOT PROPERLY PROCESSED AND THERE WAS A RED TINGE (BLOOD) TO THE COMPLETED SAMPLES. CUSTOMER IDENTIFIED THIS IS NOT CORRECT SO THEY STOPPED USING SYSTEM. CUSTOMER THEN MANUALLY PROCESSED SAMPLES BY HAND.
Additional Manufacturer Narrative
THE FOLLOWING FIELD WAS UPDATED WITH CORRECTED INFORMATION: H.6 IMDRF ANNEX G: G04135 VALVE(S). H.6 INVESTIGATION SUMMARY: BASED ON THE INVESTIGATION RESULTS, THE REPORTED ISSUE OF THE CUSTOMER EXPERIENCING SAMPLES NOT BEING PROCESSED AFTER TESTING WAS CONFIRMED. INVESTIGATION RESULTS THAT WERE PERFORMED ON THE INDICATED FAILURE MODE WERE THE FOLLOWING: THE COMPLAINT TREND AND SERVICE NOTES WERE REVIEWED. THE POTENTIAL CAUSE FOR THE CUSTOMER EXPERIENCING SAMPLES NOT BEING PROCESSED WAS LINKED TO THE 4-WAY VALVE. THE ISSUE WAS RESOLVED AFTER THE PART WAS REPLACED. THE REPLACED PART WAS NOT RETURNABLE; THEREFORE, IT WAS DISCARDED. NO FURTHER PROBLEMS WERE NOTED, AND THE INSTRUMENT WAS CONFIRMED TO BE FUNCTIONING AS EXPECTED. ALTHOUGH THE INSTRUMENT WAS USED FOR DIAGNOSTIC TESTING, THE ISSUE WAS RESOLVED BEFORE THE PATIENT SAMPLE RESULTS WERE USED FOR ANY DIAGNOSIS OR TREATMENT.
Additional Manufacturer Narrative
H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
Description of Event or Problem
IT WAS REPORTED THAT DURING USE OF BD FACS¿ SAMPLE PREP ASSISTANT III, FOREIGN MATTER/CONTAMINATION IN INSTRUMENT THAT IMPACTED PATIENT RESULTS WAS OBSERVED. CONTAMINANT IS IN A POSITION WHERE IT IMPACTED A READOUT TO BE USED CLINICALLY. THERE WAS NO PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED BY THE CUSTOMER THAT SAMPLES NOT BEING PROCESSED CORRECT. CUSTOMER REPLACED SYRINGE BUT ISSUE PERSISTS. PATIENT SAMPLES CHECKLIST: 1. DID THIS CAUSE ERRONEOUS RESULTS ON PATIENT SAMPLES FOR DIAGNOSTIC TESTING? IF YES, GO TO QUESTION #2. IF NO, NO FURTHER QUESTIONS REQUIRED. YES. 2. WAS AN ERRONEOUS LABORATORY RESULT FROM THE BDB PRODUCT REPORTED TO A CLINICIAN? IF YES, GO TO QUESTION #3. IF NO, NO FURTHER QUESTIONS REQUIRED. NO. CONTAMINATION: 1. WERE PATIENT SAMPLES CONTAMINATED OR WAS THERE CARRYOVER (CROSS-CONTAMINATION/MIXING) BETWEEN PATIENT SAMPLES? IF NO, NO FURTHER QUESTIONS REQUIRED. PATIENT SAMPLES CONTAMINATED, CARRYOVER BETWEEN PATIENT SAMPLES, UNKNOWN - GO TO QUESTION #2. PATIENT SAMPLES CONTAMINATED. 2. PLEASE DESCRIBE ANY ADDITIONAL DETAILS: GO TO PATIENT SAMPLES CHECKLIST. SAMPLES WERE NOT PROPERLY PROCESSED AND THERE WAS A RED TINGE (BLOOD) TO THE COMPLETED SAMPLES. CUSTOMER IDENTIFIED THIS IS NOT CORRECT SO THEY STOPPED USING SYSTEM. CUSTOMER THEN MANUALLY PROCESSED SAMPLES BY HAND.