Recalls for Subsystem, Water Purification
15 on file
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedJun 13, 2024
Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.
- Outset Medical, Inc.Open, ClassifiedApr 7, 2023
An observed trend of high conductivity dialysate alarms in a hemodialysis system as a result of introducing a software tool into the manufacturing process.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedSep 19, 2022
Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedSep 19, 2022
Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedJul 25, 2022
Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on the outside of the device), potentially harming service technicians and/or other authorized operators. If the ethernet port on the device is touched or manipulated, there is a risk of electrical shock
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedJul 25, 2022
Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on the outside of the device), potentially harming service technicians and/or other authorized operators. If the ethernet port on the device is touched or manipulated, there is a risk of electrical shock
- Mar Cor PurificationTerminatedMay 6, 2014
Mar Cor Purification manufactured Central Water Plant (CWP) reverse osmosis system, models 102, 104 and 106. This correction has been initiated due to a potential situation with a specific lot of the water inlet solenoid valve. The potential exists that the water inlet solenoid valve could cease to function. If this occurs, CWP will stop producing reverse osmosis water.
- Mar Cor PurificationTerminatedMar 14, 2014
High inlet water pressure beyond specification and the solenoid valve does not meet application which could result in a delay in treatment.
- Mar Cor PurificationTerminatedMar 13, 2014
There exists a potential situation with the crimp connection at the pump head assembly. The potential exists that the reverse osmosis pump head wiring could build resistance resulting in resistive heating of the surrounding components. If this occurs, the pump head assembly may experience melting of the plastic components and potentially create smoke/fire and shut down of the unit.
- AmeriWater IncTerminatedSep 6, 2012
The firm was notified that a malfunction of the device was caused by an incorrect fuse 32 volt 15 amp fuses instead of 250 volt 15 amp fuses, that was installed in the water system. The incorrect fuse may result in excessive heat in the fuse holder eventually resulting in failure of the device to operate.
- U.S. Filter/Ionpure, Inc.TerminatedMay 6, 2005
Incorrect type of stainless steel firm used 303 vs 316
- Ameriwater IncTerminatedOct 5, 2004
The device has a potential risk of failure caused by faulty wiring within the electrical component. The wires can overheat and eventually break in two.
- Fresenius Medical Care Holdings, Inc.TerminatedUnknown
Incorrectly crimped connectors were installed on the main protective earth conductor on these machines. If the machine is moved or maintenance is performed, these crimp connections could become loose and may adversely affect the electrical safety of the products, potentially resulting in electrical shock to the Biomed Technician or clinic staff.
- Fresenius Medical Care Holdings, Inc.TerminatedUnknown
AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation
- Fresenius Medical Care Holdings, Inc.TerminatedUnknown
AquaBplus HF module on the AquaBplus 2500 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2500 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation