Electrode, Ph, Stomach
Gastroenterology, Urology · 21 CFR 876.1400
Recent MDR events
12,853 adverse event reports on file
- BravoJul 31, 2026MalfunctionGiven Imaging Ltd., Yoqneam9710107-2026-00357
- BravoJul 31, 2026MalfunctionGiven Imaging Ltd., Yoqneam9710107-2026-00358
- BravoJul 31, 2026MalfunctionGiven Imaging Ltd., Yoqneam9710107-2026-00359
- BRAVOJul 21, 2026MalfunctionGiven Imaging Ltd., Yoqneam9710107-2026-00352
- BRAVOJul 21, 2026MalfunctionGiven Imaging Ltd., Yoqneam9710107-2026-00353
- BRAVOJul 21, 2026InjuryGiven Imaging Ltd., Yoqneam9710107-2026-00354
- Medtronic Bravo Reflux CapsuleJul 20, 2026OtherMW6000051
- BRAVOJul 17, 2026MalfunctionGiven Imaging Ltd., Yoqneam9710107-2026-00345
- BRAVOJul 17, 2026MalfunctionGiven Imaging Ltd., Yoqneam9710107-2026-00346
- BRAVOJul 17, 2026InjuryGiven Imaging Ltd., Yoqneam9710107-2026-00347
Event type breakdown
Recalls
10 recall events for this product code
- Given Imaging Ltd.Open, ClassifiedJul 10, 2025
Bravo CF capsule delivery device, designed to place and attach a pH monitoring capsule to the esophagus using endoscopic or manometric guidance, Product Number
It has been determined that misapplied adhesive on the Bravo CF capsule delivery device may lead to its malfunction. Specifically, the misapplied adhesive may prevent the capsule from attaching to the patient's esophagus or detaching from the delivery device.
Z-2047-2025 · View FDA record ↗ - Covidien LLCTerminatedMay 6, 2016
Given Imaging Bravo pH capsule delivery device, 5-pack.
Potential of an allergic reaction in a patient with nickel sensitivity.
Z-1619-2016 · View FDA record ↗ - Covidien LLCTerminatedMay 6, 2016
Given Imaging Bravo pH capsule delivery device, 1-pack.
Potential of an allergic reaction in a patient with nickel sensitivity.
Z-1620-2016 · View FDA record ↗ - Medtronic NeuromodulationTerminatedMay 7, 2008
Medtronic Bravo pH Capsule with Delivery System, Stomach pH electrode, Product Numbers 9012B1011 and 9012B1001. Medtronic, Inc., Minneapolis, MN 55432
Unable to detach from source: The capsule may not detach from the delivery system following attachment to the esophageal wall.
Z-0517-2008 · View FDA record ↗ - Medtronic Gastroenterology / UrologyTerminatedMay 3, 2006
Medtronic Zinetics 24ME Multi-Use External Reference pH Catheter. For Gastroesophageal pH Measurements. Catalog number 9012P2121. Distributed by Medtronic, In
Lot 0001491 has a manufacturing defect in the sensor spacing. Sensors have been placed 5 cm apart instead of 15 cm apart. Medtronic has determined that patient safety is not compromised. Clinicians however should review and reassess the results of any patient studies for which catheters from lot 0001491 were used.
Z-0830-06 · View FDA record ↗ - Medtronic Gastroenterology / UrologyTerminatedJan 13, 2006
Medtronic Slimline Single-Use Internal Reference pH Catheter for Gastroesophageal pH measurements. Models 9012P3101, 9012P3111, 9012P3121, 9012P3131 and 9012
Certain lots have a manufacturing defect that allows the water pushed through the catheter to travel back into the connector of the Digitrapper pH recorder. Medtronic has determined that patient safety is not compromised, but catheter function may be affected and damage may result to the Digitrapper pH recorder.
Z-0392-06 · View FDA record ↗ - Sandhill Scientific, IncTerminatedNov 11, 2003
ComforTEC Plus Impedance Probe, a part of the Insight S980000 Gastrointestinal motility system
Pouch containing pediatric probes was labeled as containing adult probes.
Z-0104-04 · View FDA record ↗ - Sandhill Scientific, IncTerminatedJan 23, 2003
''ComforTEC II Single Use pH Probe *** Latex Free *** Do Not Re-use''. Packaged individually.
Failure of the dry silver chloride reference, resulting in malfunction and a 'Lead Off' indication.
Z-0466-03 · View FDA record ↗ - Covidien LlcOpen, ClassifiedUnknown
CF Delivery Dev Caps Bravo x5, FGS-0635/FGS 0636 - Product Usage: intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in ad
The capsule may fail to attach to the esophageal mucosa, which in turn may lead to the aspiration of the capsule. In the event of capsule aspiration, immediate intervention to remove the capsule is required. Potential outcomes following capsule aspiration include low oxygen saturation, intervention to retrieve the capsule, potential need to intubate the patient, extended hospital stay and a delay of treatment. Thirteen reports of capsule aspiration have been received the last 2 years.
Z-1334-2021 · View FDA record ↗ - Covidien LlcOpen, ClassifiedUnknown
BRAVO PH CAPS DELIVERY DEV5-PK, FGS-0312/FGS-0313 - Product Usage: intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in a
The capsule may fail to attach to the esophageal mucosa, which in turn may lead to the aspiration of the capsule. In the event of capsule aspiration, immediate intervention to remove the capsule is required. Potential outcomes following capsule aspiration include low oxygen saturation, intervention to retrieve the capsule, potential need to intubate the patient, extended hospital stay and a delay of treatment. Thirteen reports of capsule aspiration have been received the last 2 years.
Z-1335-2021 · View FDA record ↗
Registered firms
13 firms currently registered for this product code
Chongqing Jinshan Science & Technology (group) Co., Ltd
Chongqing, CN
Given Imaging Vietnam Co., Ltd.
Ho Chi Minh City, VN
Diversatek Vietnam Company Limited
Thuan An City Binh Duong, VN
The Reflux Company
San Diego, CA, US
Diversatek Healthcare Inc.
Highlands Ranch, CO, US
Alacer Industria Eletronica Ltda
Sao Paulo, BR
Given Imaging Ltd.
Yoqneam Northern, IL
Medtronic Engineering And Innovation Center (meic)
Hyderabad Telangana, IN
510(k) clearances
49 clearances for this product code
- Given Imaging , Ltd.Dec 1, 2010
K102543 · BRAVO PH MONITORING SYSTEM AND ACCESSORIES
- Sandhill Scientific, Inc.Feb 28, 2002
K013951 · SANDHILL SCIENTIFIC PEDIATEC PH PROBE
- Endonetics, Inc.Sep 29, 2000
K002028 · BRAVO PH MONITORING SYSTEM
- Synectics-DantecJan 29, 1997
K961346 · PH ANALYSIS MODULE
- Medical Instruments Corporation Mic AGFeb 6, 1996
K950087 · LOT 440
- Instrumentarium Corp.Jan 30, 1996
K955018 · TRIP NGS & SIGMOID CATHETER PHI SLIDE CALCULATOR
- Oakfield Instruments, Ltd.Jun 13, 1995
K945069 · FLEXILOG 2000
- Tonometrics, Inc.Mar 17, 1995
K920849 · TRIP CIC CATHETER