inforMED
InjuryFFT

BRAVO

Received Jul 17, 2026 · Event occurred Jun 18, 2026

Report 9710107-2026-00347 · MDR key 25855308

Device

Generic name

Electrode, Ph, Stomach

Model number

FGS-0635

Catalog number

FGS-0635

Lot number

68248F

Product problems

  • Positioning Failure

Patient

Not reported

  • Airway Obstruction

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE CAPSULE FAILED TO ATTACH. IT WAS LOCATED IN THE PATIENT'S MOUTH AND RETRIEVED. ANOTHER CAPSULE WAS USED AND ALSO FAILED TO ATTACH. IT FELL OUT AND RETRIEVED. LUBRICATION WAS USED TO FACILITATE THE PLACEMENT OF BOTH CAPSULES. THERE WAS NO PATIENT OR USER HARM.

Additional Manufacturer Narrative

D10 CONCOMITANT PRODUCT: FGS-0635, FGS-0635 CF DELIVERY DEV CAPS BRAVO X5 (LOT#68248F). MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.