Recalls for Gauge, Pressure, Coronary, Cardiopulmonary Bypass
5 on file
- Medtronic Perfusion SystemsOpen, ClassifiedMay 10, 2024
Potential for unsealed sterile packing.
- Medtronic Perfusion SystemsTerminatedJul 10, 2017
Identification of small pinholes in a single packaging configuration of sterile pouches. A total of seven different products were affected.
- Medtronic Perfusion SystemsTerminatedOct 20, 2011
The DFU for the Medtronic Pressure Display Box, model 66000, states that, for display readings of -100 to +300 mmHg there will be an accuracy of +/-5 mmHg, and for display readings of +301 to +500 mmHg there will be an accuracy of +/-10 mmHg. Through reports from customers, Medtronic has become aware that the Pressure Display Box may not always accurately display readings. During use in the -40
- Medtronic Perfusion SystemsTerminatedOct 20, 2011
Pressure Display Set (models 61000, 61006, 62000, 62006) DFU will be updated to clarify that the intended use of the disposables is only for cardiopulmonary bypass surgery (procedures up to 6 hours). Use beyond 6 hours may result in air leakage in the system resulting in a pressure reading that is not representative of the actual line pressure.
- Medtronic Perfusion SystemsTerminatedOct 20, 2011
Pressure Display Set (models 61000, 61006, 62000, 62006) DFU will be updated to clarify that the intended use of the disposables is only for cardiopulmonary bypass surgery (procedures up to 6 hours). Use beyond 6 hours may result in air leakage in the system resulting in a pressure reading that is not representative of the actual line pressure.