inforMED
InjuryOUO

SIGNA PETMR

Received Jan 9, 2017 · Event occurred Nov 16, 2016

Report 2183553-2017-00001 · MDR key 6234046

Device

Generic name

Tomographic Imager Combining Emiss

Model number

SIGNA PETMR

Product problems

  • Adverse Event Without Identified Device or Use Problem

Patient

Not reported

  • Hearing Loss
  • Pain
  • Hearing Loss
  • Pain

Narrative

Additional Manufacturer Narrative

GE HEALTHCARE¿S INVESTIGATION IN ONGOING. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION HAS BEEN COMPLETED. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.

Description of Event or Problem

IT WAS REPORTED THAT A PATIENT UNDERGOING A PETMR EXAM COMPLAINED OF RIGHT EAR PAIN AND HEARING LOSS. THE HEARING LOSS WAS LATER CONFIRMED BY A PHYSICIAN.

Additional Manufacturer Narrative

THE INVESTIGATION BY GE HEALTHCARE HAS BEEN COMPLETED. THE SIGNA PET/MR SYSTEM LIMITS ACOUSTIC NOISE, WHEN HEARING PROTECTION IS USED. THE SYSTEM OPERATOR IS RESPONSIBLE FOR PROVIDING THE HEARING PROTECTION. IN THIS CASE, HEARING PROTECTION WAS PROVIDED AND PLACED BY THE PATIENT. THE SYSTEM ACOUSTIC LEVEL WAS TESTED TO MEET LOCAL REGULATIONS. NO SYSTEMIC ISSUE WAS FOUND. NO CORRECTIONS ARE REQUIRED AS THE SYSTEM WAS OPERATING WITHIN SPECIFICATION.