inforMED
MalfunctionMUH

Schick 3.0 USB interface

Received Jul 28, 2026

Report 2424472-2026-05830 · MDR key 50059413

Device

Generic name

System, X-ray, Extraoral Source, Digital

Catalog number

GB400

Product problems

  • Intermittent Communication Failure

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

While the customer was using a part of an intraoral sensor system, Schick 3.0 USB interface, they allege there was an x-ray image transmission issue where the x-ray image was not captured after radiation has occurred (transmission failure). No injury was reported from the alleged event.

Additional Manufacturer Narrative

While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such. This event, therefore, is reportable per 21CFR Part 803.USB 3 Firmware v1.4 contains an issue in the portion of code used to optimize sensor performance. If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted. Image data following the point of corruption will consist of data that was previously held in the memory of the USB device. As a result, the presented image may contain a portion of data from a previous exposure. In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.