SPYGLASS DS
Received Jul 20, 2026 · Event occurred Jun 30, 2026
Report 3005099803-2026-02905 · MDR key 25876865
Device
Generic name
Choledochoscope And Accessories, Flexible/rigid
Manufacturer
Boston Scientific CorporationModel number
M00546650Catalog number
M00546650
Lot number
2304700045
Product problems
- Optical Problem
- Unexpected Shutdown
Patient
Not reported
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
NOTE: THIS REPORT PERTAINS TO ONE OF THREE DEVICES USED DURING THE SAME PROCEDURE.IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SPY DS DIGITAL CONTROLLER WAS INTENDED FOR USE WITH A SPYSCOPE DS II DEVICE IN AN ENDOSCOPIC PROCEDURE ON (b)(6) 2026.DURING THE PROCEDURE, THE CONTROLLER BEGAN REBOOTING UNEXPECTEDLY AND SUBSEQUENTLY FAILED COMPLETELY, RESULTING IN LOSS OF THE VISUALIZATION. AFTER SWITCHING TO A SECOND SPYSCOPE DS II, THE SAME BEHAVIOR RECURRED. AS A RESULT, THE SYSTEM WAS NO LONGER AVAILABLE FOR THE PROCEDURE, AND THE DEVICE COULD NOT BE USED TO CONTINUE THE PROCEDURE AS INTENDED.NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Additional Manufacturer Narrative
BLOCK H6 (IMPACT CODES):IMDRF IMPACT CODE F1001 IS BEING USED TO CAPTURE THE REPORTABLE EVENT OF ABSENCE OF TREATMENT.BLOCK D2B: SUBSEQUENT PRO CODES KQM, NTN.