BRAVO
Received Jul 8, 2026 · Event occurred Jun 3, 2026
Report 9710107-2026-00322 · MDR key 25749140
Device
Generic name
Electrode, Ph, Stomach
Manufacturer
Given Imaging Ltd., YoqneamModel number
FGS-0635Catalog number
FGS-0635
Lot number
68250F
Product problems
- Positioning Failure
Patient
50 YR · Male
- Airway Obstruction
Narrative
Description of Event or Problem
IT WAS REPORTED THAT THE CAPSULE DETACHED FROM THE DELIVERY DEVICE BUT PARTIALLY ATTACHED TO ESOPHAGEAL WALL. THE PIN OF THE CAPSULE DID NOT GO ALL THE WAY THROUGH. THE CAPSULE WAS SUBSEQUENTLY LOCATED IN THE ORAL CAVITY. THE CAPSULE WAS RETRIEVED FROM THE PATIENT¿S BODY USING RAPTOR FORCEPS. ANOTHER ATTEMPT WAS MADE TO PLACE A CAPSULE, AND THE SECOND CAPSULE SUCCESSFULLY ATTACHED TO THE ESOPHAGUS. NO LUBRICATION WAS USED TO FACILITATE THE PLACEMENT OF THE CAPSULE.
Additional Manufacturer Narrative
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.