BRAVO
Received May 27, 2026 · Event occurred Apr 30, 2026
Report 9710107-2026-00255 · MDR key 25303040
Device
Generic name
Electrode, Ph, Stomach
Manufacturer
Given Imaging Ltd., YoqneamModel number
FGS-0635Catalog number
FGS-0635
Lot number
67181F
Product problems
- Positioning Failure
Patient
NA · Unknown
- Airway Obstruction
Narrative
Description of Event or Problem
IT WAS REPORTED THAT THE CAPSULE FAILED TO ATTACH TO THE PATIENT'S ESOPHAGUS AND FELL AND RETRIEVED FROM THE VOCAL CORDS. THE CAPSULE WAS RETRIEVED USING A SNARE. A SECOND CAPSULE WAS USED AND IT ALSO FELL INTO THE VOCAL CORDS AND WAS RETRIEVED USING A SNARE. NO LUBRICANT WAS USED TO FACILITATE THE PLACEMENT OF THE CAPSULE.
Additional Manufacturer Narrative
D10 CONCOMITANT PRODUCT: FGS-0635, FGS-0635 CF DELIVERY DEV CAPS BRAVO X5 (LOT#67181F) MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.