inforMED
MalfunctionFIP

ULTRAFILTER HECHINGEN

Received Aug 20, 2025 · Event occurred Jul 3, 2025

Report 9611369-2025-00122 · MDR key 22851069

Device

Generic name

Subsystem, Water Purification

Model number

NA

Catalog number

955858

Lot number

5-2150-H-01

Product problems

  • Fluid/Blood Leak
  • Fluid/Blood Leak

Patient

50 YR · Male

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

E1: INITIAL REPORTER ADDRESS: (B)(6). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem

IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK WAS OBSERVED FROM AN ULTRAFILTER U9000 DURING HEMODIALYSIS THERAPY. THE PATIENT'S WEIGHT WAS 0.2KG MORE THAN THE PRESET AMOUNT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Additional Manufacturer Narrative

ADDITIONAL INFORMATION: H6, H11. H11: A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. THE DEVICE WAS NOT RECEIVED FOR EVALUATION; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.