inforMED
MalfunctionFIP

ULTRAFILTER HECHINGEN

Received May 21, 2025 · Event occurred Apr 29, 2025

Report 9611369-2025-00075 · MDR key 22062638

Device

Generic name

Subsystem, Water Purification

Model number

NA

Catalog number

114604

Lot number

4-1805-H-01

Product problems

  • Fluid/Blood Leak
  • Fluid/Blood Leak

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK WAS OBSERVED FROM AN ULTRAFILTER U9000 "JUST AFTER INSTALLATION" IN AN AK98 MACHINE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Additional Manufacturer Narrative

SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative

ADDITIONAL INFORMATION: H6, H11 H11: THE ACTUAL DEVICE WAS NOT AVAILABLE; HOWEVER, PHOTOGRAPHS OF THE SAMPLE WERE PROVIDED FOR EVALUATION. VISUAL INSPECTION OF THE PHOTOS DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED LEAKAGE. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE WAS NOT DETERMINED AS NO FURTHER OR ACTUAL SAMPLE TESTING COULD BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.