Subsystem, Water Purification

Manufacturer

Vivonic Gmbh

Catalog number

24-180A-1

Product problems

  • Thermal Decomposition of Device
  • Thermal Decomposition of Device

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

A USER FACILITY BIOMEDICAL TECHNICIAN (BIOMED) REPORTED TO FRESENIUS TECHNICAL SERVICES THAT ERROR "W¿02¿02¿17 WARNING: STAGE 2, CONC. PRESSURE BELOW ALARM LIMIT" WAS RECEIVED IN SUPPLY MODE. THERMAL DECOMPOSITION WAS IDENTIFIED UPON EVALUATION. THE OUTLET SOCKET OF THE POWER DISTRIBUTION CENTER WAS CHARRED AS WAS THE POWER PLUG THAT FEEDS THE RO SYSTEM. ADDITIONAL INFORMATION WAS OBTAINED DURING FOLLOW-UP. A LIGHT BURNING SMELL WAS NOTED, BUT NO SMOKE, SPARK, FLAME, OR ARCING WAS OBSERVED. NO BLOWN FUSES WERE IDENTIFIED. THERE WERE NO LOCAL POWER GRID ISSUES OR ELECTRICAL STORMS IN THE AREA ON OR AROUND THE REPORTED EVENT DATE. THERE WAS NO PATIENT INVOLVEMENT NOR PERSONAL HARM TO ANY PATIENTS OR INDIVIDUALS AS A RESULT OF THE REPORTED ISSUE. THE PLUG AND SOCKET WERE REPLACED TO RESOLVE THE REPORTED ISSUE. THE RO SYSTEM HAS BEEN RETURNED TO SERVICE. NO PARTS WERE RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION.

Additional Manufacturer Narrative

THE PLANT INVESTIGATION IS IN PROCESS. A SUPPLEMENTAL MDR WILL BE SUBMITTED UPON COMPLETION OF THIS ACTIVITY.