7052HP WASHER/DISINFECTOR
Received May 17, 2024 · Event occurred Apr 17, 2024
Report 9680353-2024-00019 · MDR key 19340665
Device
Generic name
Washer/disinfector
Manufacturer
Steris Canada UlcProduct problems
- Leak/Splash
- Leak/Splash
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
A STERIS SERVICE TECHNICIAN ARRIVED ONSITE TO INSPECT THE AMSCO 7052HP WASHER AND FOUND ONE OF THE TRI-CLAMP FITTINGS BECAME LOOSE AND DETACHED FROM THE RECIRCULATION PIPE CAUSING THE SUMP PUMP TO LEAK OUT ON THE FLOOR. THE TECHNICIAN SECURED AND TIGHTENED THE HARDWARE. THE TECHNICIAN TESTED THE UNIT AND CONFIRMED TO BE OPERATING TO SPECIFICATION AND RETURNED TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.
Description of Event or Problem
THE USER FACILITY REPORTED THAT THEIR AMSCO 7052HP WASHER LEAKED WATER IN THE AREA SURROUNDING THE UNIT. NO REPORT OF INJURY OR PROCEDURE DELAY.