inforMED
MalfunctionFIP

AQUABPLUS, HF 2000

Received Oct 25, 2023 · Event occurred Aug 1, 2021

Report 3010850471-2023-00090 · MDR key 18006172

Device

Generic name

Subsystem, Water Purification

Manufacturer

Vivonic Gmbh

Catalog number

24-2000-1

Lot number

N/A

Product problems

  • Thermal Decomposition of Device
  • Thermal Decomposition of Device

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

PLANT INVESTIGATION: THE COMPLAINT INVESTIGATION DETERMINED A BAD ELECTRICAL CONTACT OF THE WIRING AT THE MOTOR PROTECTION SWITCH AS THE FAILURE CAUSE. IN THE CURRENT DESIGN THE MOTOR PROTECTION SWITCH IS ALSO THE MAIN SWITCH OF THE DEVICE. THE DESIGN OF THE BLADE RECEPTACLE AT THE MOTOR PROTECTION SWITCH IS INADEQUATE. THE BAD ELECTRICAL CONTACT AT THE MOTOR PROTECTION SWITCH RESULTS IN THE PUMP P1(S) RUNNING ONLY WITH TWO LINE CONDUCTORS AND RESULTING IN HIGHER CURRENT AND HIGHER THERMAL ENERGY. OR THE INNER BIMETAL CONTACT OF THE MOTOR PROTECTION DO TRIP DUE TO INCREASED TEMPERATURE CAUSED BY THE TRANSITION RESISTANCE OF THE ELECTRICAL CONTACT. IN BOTH CASES THE MOTOR PROTECTION SWITCH TRIPS AND THE DEVICE IS POWERED OFF. THIS IS A KNOWN FAILURE. A ROOT CAUSE ANALYSIS IS IN PROGRESS AND CORRECTIVE/PREVENTIVE ACTIONS WILL BECOME AVAILABLE WITH AN IMPROVED DESIGN FOR THE ELECTRICAL MONITOR, WHICH INCLUDES THE WIRING AT THE MAIN. THE IMPROVED DESIGN IS CURRENTLY NOT AVAILABLE FOR THE AFFECTED MARKET SEGMENT, BUT THE RELEASE IS PLANNED BY COMPLETION OF THE DESIGN CHANGE IMPLEMENTATION. BASED ON THE AVAILABLE PHOTOGRAPHS, LOCAL PARTS WERE USED TO REPAIR THE MACHINE AND RESOLVE THE PROBLEM. ONLY ORIGINAL SPARE PARTS ARE ADVISED FOR USE PER THE SERVICE MANUAL AND INSTRUCTIONS FOR USE (IFU). THE CURRENT STATUS OF THE MACHINE IS UNKNOWN. IT IS RECOMMENDED TO REPLACE, IF NOT ALREADY DONE, THE WIRING AND THE MOTOR PROTECTION SWITCH IN STAGE 1 AND STAGE 2 WITH THE AVAILABLE DESIGN.

Description of Event or Problem

IT WAS REPORTED TO FRESENIUS THAT THERMAL DAMAGE WAS IDENTIFIED WITHIN THE AQUABPLUS REVERSE OSMOSIS (RO) SYSTEM. THE AFFECTED COMPONENTS WERE ISOLATED TO THE IGNITION SWITCH GROUP, ELECTRICAL CONTACTOR OF THE PRESSURE PUMP, AND ELECTRICAL WIRING OF THE CONNECTION BETWEEN THE MAIN IGNITION SWITCH AND THE ELECTRICAL CONTACTOR THAT FEEDS THE PRESSURE PUMP OF THE AQUABPLUS EQUIPMENT. IT WAS REPORTED THAT THE IGNITION SWITCH CABLE FAILED. THE MACHINE HAS APPROXIMATELY11,881 OPERATING HOURS. THE CABLE WAS REPLACED. ACCORDING TO THE PROVIDED REPAIR INFORMATION, TERMINALS AND FAULTY CONTACTORS WERE ALSO REPLACED TO RESOLVE THE REPORTED DAMAGE, IN ADDITION TO THE CABLE GAUGE TO ALLOW FOR A GREATER LOAD OF CURRENT INTENSITY. IT WAS NOT REPORTED WHETHER A SAMPLE WAS AVAILABLE TO BE RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION. THERE WAS NO PATIENT INVOLVEMENT NOR REPORTED PERSONAL HARM TO ANY INDIVIDUAL AS A RESULT OF ENCOUNTERING THE THERMAL DAMAGE.

Description of Event or Problem

IT WAS REPORTED TO FRESENIUS THAT THERMAL DAMAGE WAS IDENTIFIED WITHIN THE AQUABPLUS REVERSE OSMOSIS (RO) SYSTEM. THE AFFECTED COMPONENTS WERE ISOLATED TO THE IGNITION SWITCH GROUP, ELECTRICAL CONTACTOR OF THE PRESSURE PUMP, AND ELECTRICAL WIRING OF THE CONNECTION BETWEEN THE MAIN IGNITION SWITCH AND THE ELECTRICAL CONTACTOR THAT FEEDS THE PRESSURE PUMP OF THE AQUABPLUS EQUIPMENT. IT WAS REPORTED THAT THE IGNITION SWITCH CABLE FAILED. THE MACHINE HAS APPROXIMATELY 11,881 OPERATING HOURS. THE CABLE WAS REPLACED. ACCORDING TO THE PROVIDED REPAIR INFORMATION, TERMINALS AND FAULTY CONTACTORS WERE ALSO REPLACED TO RESOLVE THE REPORTED DAMAGE, IN ADDITION TO THE CABLE GAUGE TO ALLOW FOR A GREATER LOAD OF CURRENT INTENSITY. IT WAS NOT REPORTED WHETHER A SAMPLE WAS AVAILABLE TO BE RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION. THERE WAS NO PATIENT INVOLVEMENT NOR REPORTED PERSONAL HARM TO ANY INDIVIDUAL AS A RESULT OF ENCOUNTERING THE THERMAL DAMAGE.

Additional Manufacturer Narrative

THE PLANT INVESTIGATION IS IN PROCESS. A SUPPLEMENTAL MDR WILL BE SUBMITTED UPON COMPLETION OF THIS ACTIVITY.