inforMED
MalfunctionPER

BD FACS¿ SAMPLE PREP ASSISTANT

Received Mar 21, 2022 · Event occurred Mar 10, 2022

Report 2916837-2022-00063 · MDR key 13844312

Device

Generic name

See H10

Catalog number

650686

Lot number

NA

Product problems

  • Fluid/Blood Leak
  • Fluid/Blood Leak

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT DURING USE WITH BD FACS¿ SAMPLE PREP ASSISTANT THE WAS TOWER WAS OVERFLOWING AND BIOHAZARDOUS WASTE LEAKED OUTSIDE OF INSTRUMENT. THERE WAS NO USER IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THE WASH TOWER IS OVER FLOWING. WAS THE LEAK LIQUID OR AIR? LIQUID. WAS THE LEAK CONTAINED WITHIN THE INSTRUMENT? NOT CONTAINED. WAS THERE SPRAY OF LIQUID? NO. WHAT WAS THE FLUID THAT LEAKED? BIOHAZARD. DID BIOHAZARD LEAK BEFORE OR AFTER WASTE LINE? BEFORE WASTE LINE. WAS THE CUSTOMER/BD PERSONNEL PHYSICALLY IN CONTACT WITH THE FLUID? NO.

Additional Manufacturer Narrative

COMMON DEVICE NAME: AUTOMATED PIPETTING, DILUTING AND SPECIMEN PROCESSING WORKSTATIONS FOR FLOW CYTOMETRIC ANALYSIS. MEDICAL DEVICE EXPIRATION DATE: NA. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem

IT WAS REPORTED THAT DURING USE WITH BD FACS¿ SAMPLE PREP ASSISTANT THE WAS TOWER WAS OVERFLOWING AND BIOHAZARDOUS WASTE LEAKED OUTSIDE OF INSTRUMENT. THERE WAS NO USER IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THE WASH TOWER IS OVER FLOWING. WAS THE LEAK LIQUID OR AIR? LIQUID. WAS THE LEAK CONTAINED WITHIN THE INSTRUMENT? NOT CONTAINED. WAS THERE SPRAY OF LIQUID? NO. WHAT WAS THE FLUID THAT LEAKED? BIOHAZARD. DID BIOHAZARD LEAK BEFORE OR AFTER WASTE LINE? BEFORE WASTE LINE. WAS THE CUSTOMER/BD PERSONNEL PHYSICALLY IN CONTACT WITH THE FLUID? NO.

Additional Manufacturer Narrative

H.6. INVESTIGATION: SCOPE OF ISSUE: THE SCOPE OF ISSUE IS LIMITED TO PART: 650686 SPAIII AND SERIAL NUMBER: (B)(6). PROBLEM STATEMENT: CUSTOMER REPORTED: WASH STATION IS OVERFLOWING. MANUFACTURING DEFECT TREND: THERE ARE 0 QNS RELATED TO THE REPORTED ISSUE. DATE RANGE (DATE OF INCIDENT TO 12 MONTHS BACK) FROM 10MAR2021 TO DATE 10MAR2022 (ROLLING 12 MONTHS). COMPLAINT TREND: THERE ARE 9 COMPLAINTS RELATED TO THE REPORTED COMPLAINT. DATE RANGE (DATE OF INCIDENT TO 12 MONTHS BACK) FROM 10MAR2021 TO DATE 10MAR2022 (ROLLING 12 MONTHS). INVESTIGATION RESULT / ANALYSIS: PER FSE REPORT: THE CUSTOMER REPORTED THAT THE WASH STATION WAS OVERFLOWING. FOUND POOR ASPIRATION FROM WASTE PUMP. OPENED WASTE PUMP AND CLEANED DEBRIS OFF FLAPPERS. WASTE PRESSURE PROPER AFTER CLEANING AND STATION DID NOT OVERFLOW. RAN SAMPLE AND LYSE DISPENSE-PASSED. THE INSTRUMENT WAS TESTED AND VERIFIED AS OPERATING TO SPECS. THE PROBLEM WAS CORRECTED AND THERE WAS NO FURTHER LEAKING OR BLOCKAGE. SAMPLE PREPARATION WAS NOT AFFECTED. THERE WAS NO DELAY IN PATIENT TREATMENT DUE TO ANY UNEXPECTED RESULTS. SERVICE MAX REVIEW: REVIEW OF RELATED WORK ORDER# (B)(4). INSTALL DATE: (B)(6) 2006. DEFECTIVE PART NUMBER: THERE WERE NO DEFECTIVE PARTS WORK ORDER NOTES: SUBJECT / REPORTED: WASH STATION WAS OVERFLOWING, PROBLEM DESCRIPTION: WASH STATION PUMP IS CLOGGED, CAUSE: CLOGGED PUMP, WORK PERFORMED: CLEARED THE CLOGGED PUMP, SOLUTION: CLEARED THE CLOGGED PUMP. RETURNED SAMPLE EVALUATION: THERE WERE NO DEFECTIVE PARTS. MANUFACTURING DEVICE HISTORY RECORD (DHR) REVIEW: REVIEW OF THE DHR FOR SERIAL NUMBER: (B)(6) AND (B)(4) WAS REVIEWED. THE INSTRUMENT MET ALL THE MANUFACTURING SPECIFICATIONS PRIOR TO RELEASE. RISK ANALYSIS: RISK MANAGEMENT FILE PART #100245RA, REVISION 03 WAS REVIEWED. HAZARD(S) IDENTIFIED? YES, NO. HAZARD ID: (B)(6), HAZARD: ENVIRONMENTAL BIOHAZARD, SEVERITY: 5, PROBABILITY: 1, RISK INDEX: 5, IMPLEMENTATION: BD FACS SAMPLE PREP USER¿S GUIDE, RISK CONTROL:_AFAP, MITIGATION(S) SUFFICIENT YES , NO. ROOT CAUSE: BASED ON THE INVESTIGATION RESULT AND THE FSE¿S REPORT THE ROOT CAUSE WAS A CLOGGED WASTE PUMP. CONCLUSION: BASED ON THE INVESTIGATION RESULTS AND THE FSE¿S REPORT THE COMPLAINT WAS CONFIRMED FOR THE WASH STATION OVERFLOWING. NO ONE WAS HARMED OR INJURED, AND NO MEDICAL DIAGNOSIS WAS PERFORMED DUE TO ANY INCORRECT RESULTS. THE SAFETY RISK IS LIMITED, S2, AND THERE WAS NO IMPACT TO THE PATIENT HEALTH OR SAFETY. H3 OTHER TEXT : SEE H.10.