Test, Carbohydrate Deficient Transferrin
Clinical Chemistry · 21 CFR 862.1360
Recent MDR events
0 adverse event reports on file
No MDR events on file for this product code.
Event type breakdown
No data available.
Most-reported manufacturers
No data available.
Recalls
2 recall events for this product code
- CSL Behring GmbHTerminatedUnknown
Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum.
Negative bias for the carbohydrate-deficient transferrin (CDT) measurement in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring
Z-1790-2020 · View FDA record ↗ - CSL Behring GmbHTerminatedUnknown
N Latex CDT Kit
Siemens Healthcare Diagnostics has observed a negative bias for the carbohydrate-deficient transferrin (CDT) measurement when using the affected N Latex CDT lots in comparison to the HPLC method. CDT results in absolute concentrations may be influenced by patients transferrin levels and, therefore, results are reported as ratio of CDT to total transferrin, called %CDT. As the transferrin determination with N Antiserum to Human Transferrin is not affected, the observed effect leads to calculated %CDT values that show a negative bias of approximately 15% compared to the HPLC method. This could result in a shift of weak positive patient samples into the reference range of 1.19 2.47 %CDT which was derived from a study population of healthy adults. There are no control or calibrator materials that would catch this bias.
Z-2931-2018 · View FDA record ↗
Registered firms