Recalls for Assay, Hybridization And/or Nucleic Acid Amplification For Detection Of Hepatitis C Rna,hepatitis C Virus
11 on file
- Abbott Molecular, Inc.Open, ClassifiedNov 29, 2023
Abbott has received reports of an increase in reactive negative controls and false positive results in patient samples when using the Identified Lots of Alinity m HCV AMP Kit.
- Roche Molecular Systems, Inc.TerminatedJun 17, 2011
The COBAS Ampliprep/COBAS TaqMan HCV Test has been shown to under-quantitate a subset of genotype 4 patient specimens by approximately 1.0-1.5 log 10 in the absence of any sequence mismatches.
- Roche Molecular Systems, Inc.TerminatedJun 17, 2011
The COBAS Ampliprep/COBAS TaqMan HCV Test has been shown to under-quantitate a subset of genotype 4 patient specimens by approximately 1.0-1.5 log 10 in the absence of any sequence mismatches.
- Roche Molecular Systems, Inc.TerminatedAug 2, 2007
For select COBAS AMPLICOR Tests run on the COBAS AMPLICOR Analyzer in conjunction with AMPLILINK software, a discrepancy has been identified between the onboard working reagent stability information reported by AMPLILINK Software (versions 1.1, 1.3, 1.4 and 2.41) and the stability information provided in the test kit package inserts/method manual.
- Roche Molecular Systems IncTerminatedAug 19, 2004
An increased frequency of 'blue foci' that potentially can cause elevated A450 background in microwell plate wells after the addition of conjugate reagent during PCR detection.
- Roche Molecular Systems IncTerminatedAug 19, 2004
An increased frequency of 'blue foci' that potentially can cause elevated A450 background in microwell plate wells after the addition of conjugate reagent during PCR detection.
- Abbott Molecular, Inc.Open, ClassifiedUnknown
There is a potential for abnormal (non-sigmoidal) amplification curves.
- Abbott Molecular, Inc.Open, ClassifiedUnknown
There is a software issue associated with the ability to properly complete a Field Service Engineer Maintenance and Diagnostics procedure.
- Abbott Molecular, Inc.Open, ClassifiedUnknown
There is a software defect that can cause the Amplification Detection clamp CLEAN position to be set too high.
- Abbott Molecular, Inc.TerminatedUnknown
There is an issue with the installation of updated camera firmware on the system.
- Abbott Molecular, Inc.CompletedUnknown
Incorrect optical calibration