Recalls for Enzyme Immunoassay, Tracrolimus
12 on file
- Siemens Healthcare Diagnostics, Inc.Open, ClassifiedJun 10, 2026
A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.
- Siemens Healthcare Diagnostics, Inc.Open, ClassifiedMar 15, 2023
Siemens Healthineers has received customer complaints and confirmed imprecision for Quality Control (QC) and patient samples with Dimension Tacrolimus (TAC) lots GA2286, GA3047 and GA3171. The health consequences of erroneous patient results are major, since in a worst case scenario, a patient s tacrolimus dosing could be altered, resulting in potential organ failure and rejection (due to too low or no tacrolimus being administered) or toxicity (due to too much tacrolimus being administered). Siemens internal investigation of the Dimension TAC assay showed the worst case imprecision with patient samples at the low end of the Analytical Measurement Range (AMR). A patient sample at 2.0 ng/mL (2.6 nmol/L) recovered as 0.0 ng/mL (0.0 nmol/L) (100% negative bias), another patient sample at 2.1 ng/mL (2.7 nmol/L) recovered at 4.3 ng/mL (5.6 nmol/L) (103% positive bias).
- Diasorin Inc.TerminatedApr 20, 2015
Instability in the PRO-Trac II Tacrolimus ELISA kit which may cause falsely elevated patient results.
- Siemens Healthcare Diagnostics, Inc.TerminatedDec 16, 2013
Siemens has confirmed that the TACR method may demonstrate reduced on-board stability which may result in imprecise and inaccurate QC and patient results.
- Siemens Healthcare Diagnostics, Inc.TerminatedAug 7, 2013
Siemens initiated a recall due to confirmed complaints of low patient sample recovery with Dimension TACR Flex Reagent cartridge (DF107) lot BB4087 and its linked calibrator (DC 107) lot 3BD029. Falsely low Tacrolimus results across the assay range may lead to alterations in dosing.
- Siemens Healthcare Diagnostics, Inc.TerminatedMar 15, 2013
Siemens has confirmed customer complaints of low patient sample recovery with Dimension(R) TACR Flex(R) lot FA3085. Quality Control (QC) materials have not exhibited low recovery of the same magnitude. Internal method regression testing versus LC/MS has shown an average 25% low bias compared to LC/MS across the assay range. Individual patients, particularly samples with lower TACR values
- Abbott LaboratoriesTerminatedOct 15, 2010
There are reagent barcode read errors when using reagent lot 86599M500 due to the print quality of the barcode.
- Siemens Healthcare Diagnostics, Inc.TerminatedSep 13, 2010
Low patient sample recovery results in all affected lots, with possibility of inappropriate increase in tacrolimus dosage based on false low result.
- Diasorin Inc.TerminatedApr 13, 2005
The software program written for use in the automated microtiter plate reader (Bio-Tek ELx800) for use with the Pro-Trac II Tacrolimus ELISA kit was found to be faulty. Assay software program did not include the Calibrator Zero OD specification (STDO>=1.500) as specified in the product insert.
- Abbott Laboratories HPD/ADD/GPRDTerminatedDec 17, 2003
Inadequate instructions for use resulting in calibration errors and the protein interference data in the package insert is not representative of assay performances.
- Microgenics CorporationOpen, ClassifiedUnknown
Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay.
- Siemens Healthcare Diagnostics, Inc.TerminatedUnknown
Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results.