Recalls for Assay, Glycosylated Hemoglobin
103 on file
- GET TESTED INTERNATIONAL ABOpen, ClassifiedDec 1, 2025
Distribution without premarket approval/clearance.
- Mckesson Medical-Surgical Inc. Corporate OfficeOpen, ClassifiedMay 27, 2025
transit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failure could increase the probability of an incorrect test result or a delayed test result.
- Polymer Technology Systems, Inc. dba PTS Diagnostics, Inc.CompletedOct 24, 2024
The reason for this recall is PTS Diagnostics, Inc. received complaints of non-matching cartridge and analyzer codes within the product test system box. Use of cartridges and analyzers with non-matching codes may result in inaccurate test results as components within each lot number are meant to be used together, and not mixed between lots. Therefore, the test results may possibly lead to Hemoglobin A1C positive or negative bias results. While the product is for monitoring only, over recovery with a higher than accurate value can be an indication for increased therapy; such as physical activity, weight reduction, meal management, pharmaceutical therapy.
- Abbott Diagnostics Technologies ASOpen, ClassifiedAug 27, 2024
Five Afinion 2 (AF2) Analyzer were incorrectly market configured as Moderate Complexity and distributed to customers. Use of moderately complex assays by waived laboratories could result in erroneous results.
- Siemens Healthcare Diagnostics, Inc.Open, ClassifiedMar 19, 2024
The reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test results could be reported and potentially lead to incorrect patient management of glucose control. A positive bias may be observed with eleven HbA1c affected lots. Positive bias was observed when the performance of these five HbA1c lots was compared to NGSP pooled patient target-value assigned samples and the maximum bias observed was 19%.
- Siemens Healthcare Diagnostics, Inc.TerminatedJun 5, 2020
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
- Siemens Healthcare Diagnostics, Inc.TerminatedJun 5, 2020
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
- Roche Diagnostics Operations, Inc.TerminatedNov 11, 2016
Roche Diagnostics has confirmed elevated QC and patient sample recovery for individual Tina-quant Hemoglobin A1c Gen.2 and Tina-quant Hemoglobin A1cDx Gen.2 cassettes/bottles. This issue occurred over multiple lots. Single cassettes/bottles of the reagents are affected. The issue can lead to erroneously high HbA1c results. If a patient with known diabetes is affected, an erroneously high HbA1c result might lead to therapeutic consequences, such as therapy escalation (introduction of another oral antidiabetic medication or insulin) or increase of the dosage. This can further lead to an increased risk of hypoglycemia. Considering the unreliable detectability of the issue, relevant medical risk for the patient cannot entirely be excluded.
- Bio-Rad Laboratories, Inc.TerminatedJan 27, 2016
On a rare occasion, there is a potential to assign a patient result to an incorrect sample ID when running in D10 Rack Loader configuration.
- Siemens Healthcare Diagnostics, Inc.TerminatedNov 25, 2015
Reagent in these lots may demonstrate an increased occurrence of high %HbA1c bias, a positive bias averaging 0.6% HbA1c units, ranging from -0.1% to 1.1% HbA1c units. The maximum bias was observed at higher %HbA1c concentrations. QC samples may exhibit a similar bias.
- Tosoh Bioscience, Inc.TerminatedJul 14, 2015
The G7 Variant Elution Buffer HiS no. 1 (M) is incorrectly labeled with an expiration date of October 2015. The correct expiration date is October 2014.
- Siemens Healthcare Diagnostics, Inc.TerminatedNov 21, 2014
Certain lots of the reagent cartridges show confirmed positive bias averaging 0.4% Hemoglobin A1c units and occasionally up to 1.0% HbA1c units for patient samples when compared to the National Glycohemoglobin Standardization Program (NGSP). QC samples may exhibit a similar bias.
- Siemens Healthcare Diagnostics, Inc.TerminatedOct 24, 2014
In March, Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction for decreased recovery of patient samples with the Dimension Vista® CSA & CSAE Flex® reagent cartridges when whole blood samples are run from Small Sample Containers (SSC). Siemens has confirmed that when whole blood samples are run from SSC, HbA1c values may exhibit a percent bias ranging from - 8.2% t
- Bio-Rad Laboratories, Inc.TerminatedMar 27, 2014
On a rare occasion the D-10 software might generate an extra result by duplicating the result form the previous sample and the Sample ID/Injection number from the next sample.
- Bio-Rad Laboratories, Inc.TerminatedFeb 10, 2014
D-10 A1c Dual Program Floppy was programmed incorrectly and cartridge injections counts are not being updated correctly when switching between the D-10 Dual Short Program and Extended Program on the D-10 Hemoglobin Testing System.
- Siemens Healthcare Diagnostics, Inc.TerminatedJan 3, 2014
Positive bias on the lots listed compared to alternate methods and a higher frequency of "Above Assay Range" flags with the lots listed.
- Siemens Healthcare Diagnostics, Inc.TerminatedJan 3, 2014
Positive bias on the lots listed compared to alternate methods and a higher frequency of "Above Assay Range" flags with the lots listed.
- Bayer Healthcare LlcTerminatedNov 2, 2013
Test results from these lots may produce a high-bias reading (by up to 20%) that is outside of limits defined for stability and product accuracy when product is stored at room temperature four (4) to five (5) months after release.
- Siemens Healthcare DiagnosticsTerminatedSep 3, 2013
Siemens confirmed that the ADVIA Chemistry Systems HbA1c method exhibits a positive bias of up to 12% for patient and CAP survey samples due to over-recovery when using A1c_3 Calibrator lots 1MD014, 1BD063, and 2GD014. The established ranges of CAP Survey Samples GH2-04, GH2-05 and GH2-06 and IFCC sample recoveries use 7% from the established mean as their acceptance criteria. This is the criterion that was used during testing to confirm the customer complaints. Depending upon quality control limits this issue may not have been detected..
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Heritage Labs International LLCTerminatedJan 11, 2012
Kits contain recalled Triad alcohol pads
- Bio-Rad Laboratories, Inc.TerminatedOct 7, 2011
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a definitive root cause and corrective action have been found and implemented.
- Bio-Rad Laboratories, Inc.TerminatedOct 7, 2011
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a definitive root cause and corrective action have been found and implemented.
- Bio-Rad Laboratories, Inc.TerminatedOct 7, 2011
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a definitive root cause and corrective action have been found and implemented.
- Bio-Rad Laboratories, Inc.TerminatedOct 7, 2011
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a definitive root cause and corrective action have been found and implemented.
- Bio-Rad Laboratories, Inc.TerminatedOct 7, 2011
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a definitive root cause and corrective action have been found and implemented.
- Bio-Rad Laboratories, Inc.TerminatedOct 7, 2011
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a definitive root cause and corrective action have been found and implemented.
- Bio-Rad Laboratories, Inc.TerminatedOct 7, 2011
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a definitive root cause and corrective action have been found and implemented.
- Bio-Rad Laboratories, Inc.TerminatedOct 7, 2011
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a definitive root cause and corrective action have been found and implemented.
- Bio-Rad Laboratories, Inc.TerminatedOct 7, 2011
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a definitive root cause and corrective action have been found and implemented.
- Bio-Rad Laboratories, Inc.TerminatedOct 7, 2011
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a definitive root cause and corrective action have been found and implemented.
- Bio-Rad Laboratories, Inc.TerminatedOct 7, 2011
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a definitive root cause and corrective action have been found and implemented.
- Bio-Rad Laboratories, Inc.TerminatedSep 6, 2011
The Elution Buffer B, lot no. GA 10430, was shipped in a 2.0 L bottle instead of the standard 2.5 L bottle.
- Bio-Rad Laboratories IncTerminatedJul 27, 2011
Reports of calibration failure and late retention time using Elution Buffer AA10434 and AA10435.
- Beckman Coulter Inc.TerminatedMay 10, 2011
The recall was initiated because Beckman Coulter has identified a problem with two lots of HbA1c APT reagent OSR61177. When HbA1c APT reagent lots 9390 and 9472 are calibrated with HbA1c calibrator ODR3032, elevated recovery may be observed for patient samples. Falsely elevated HbA1c values may be interpreted as increased risk for diabetes or poorly controlled diabetes. Consequent medical acti
- Bio-Rad Laboratories, Inc., Hercules, CATerminatedApr 13, 2011
The firm discovered that the Update Kit Floppy Diskette included in the reorder pack was mislabeled with an incorrect Lot Number for the Calibrator/Diluent Set as: S01819; The correct Lot Number for the Calibrator/Diluent Set is: S01704.
- Bio-Rad Laboratories IncTerminatedApr 11, 2011
Some customers have reported the occurrence of some ramping baselines on the chromatograms. The ramping baseline can affect hemoglobin test results.
- Bio-Rad Laboratories IncTerminatedApr 6, 2011
Wrong lot number was found on the Hemoglobin Cartridge Insert . The insert did not match the lot number to the device cartridge or the device cartridge label or on the Update Kit CD.
- Bio-Rad Laboratories IncTerminatedFeb 17, 2011
The firm received customer complaints associated with the Update Kit CD-ROM not being able to generate a valid calibration after being uploaded, resulting in patient results not being generated.
- Bio-Rad Laboratories IncTerminatedFeb 15, 2011
The results from the blood device had the occurrence of some ramping baselines on their chromatograms. The ramping baseline can affect quantitation of the blood results.
- Bio-Rad Laboratories IncTerminatedOct 22, 2010
Inaccurate results with high bias-- Inaccurate results may be read by the machine, creating the risk for untreated hypoglycemia at elevated altitudes.
- Bio-Rad Laboratories IncTerminatedOct 21, 2010
Test Cartridges may leak and have the potential to generate error messages and biased inaccurate results.
- Bio-Rad Laboratories IncTerminatedJun 25, 2010
Sample test results for the first run appears to be correct but all subsequent results during that run were assigned incorrect barcode and tube position.
- Bio-Rad Laboratories IncTerminatedJun 25, 2010
Sample test results for the first run appears to be correct but all subsequent results during that run were assigned incorrect barcode and tube position.
- Bio-Rad Laboratories IncTerminatedMar 5, 2010
Calibration Failure-- due to high intercept values, the device fails calibration and users are unable to run samples.
- Bio-Rad Laboratories IncTerminatedFeb 26, 2010
Product update kit CD rom will not upload. As a result, the device cannot be run.
- Bio-Rad Laboratories IncTerminatedMay 19, 2009
The wrong barcode number and tube position number may be assigned to samples leading to discrepant sample results.
- Bio-Rad Laboratories IncTerminatedMar 24, 2009
Sample results may be assigned to the wrong barcode number and tube position numbers, leading to discrepant sample results.
- Alfa Wassermann, Inc.TerminatedDec 17, 2008
Internal studies observed occasional unexpected outlier results in whole blood samples tested for Hemoglobin A1c using the EZA1c reagent on the ACE or ACE Alera clinical chemistry systems.
- Cholestech CorpTerminatedSep 23, 2008
Cholestech Corp. was notified by the foreign manufacturer of the A1C test cartridge that stability data indicated that the product would not meet performance claims through the end of its shelf life.
- Provalis DiagnosticsTerminatedNov 18, 2005
Provalis Diagnostics discovered, following its internal stability testing program (mid-point batch testing), that occasional low values were being found across multiple lots of in2it HbA1C Test Cartridges.
- Roche Diagnostics Corp.TerminatedMar 23, 2005
The product is not uniform across the cassettes of some lots, which will result in an inconsistent bias of patient results.
- Roche Diagnostics Corp.TerminatedMar 23, 2005
The product is not uniform across the cassettes of some lots, which will result in an inconsistent bias of patient results.
- Roche Diagnostics Corp.TerminatedMar 23, 2005
The product is not uniform across the cassettes of some lots, which will result in an inconsistent bias of patient results.
- Randox Laboratories Ltd.TerminatedUnknown
Randox confirmed low calibration absorbance with HbA1c, Catalog Number HA3830, Batch 414553. The reagent may not continue to meet its specific performance characteristics throughout the shelf life of the product.
- Metrika IncTerminatedUnknown
Incorrect Test Results: Due to manufacturing errors regarding temperature compensation factors, the device will give erroneously high results if the test is performed below 70 degrees Fahrenheit.
- Tosoh Bioscience IncTerminatedUnknown
Tosoh Bioscience has become aware of potential degradation of HbA1c control lot # 7055 before the 60-days product stability claim. If degradation of these controls occurs, P00 peaks may be seen and the % HbA1c control values may drift higher than the assigned range. This may result in delays when performing routine instrument quality control. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received two (2) complaints related to this issue with no serious injuries reported.
- Polymer Technology SystemsTerminatedUnknown
A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
- Polymer Technology SystemsTerminatedUnknown
A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
- Polymer Technology SystemsTerminatedUnknown
A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
- Polymer Technology SystemsTerminatedUnknown
A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
- Polymer Technology SystemsTerminatedUnknown
A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
- Siemens Healthcare Diagnostics, Inc.Open, ClassifiedUnknown
Siemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results.
- Tosoh Bioscience IncTerminatedUnknown
The assay can potentially generate erroneously elevated or erroneously decreased HbA1c patient results.
- Siemens Healthcare Diagnostics, Inc.TerminatedUnknown
Siemens identified an issue with the ratio equation provided on the ADVIA Chemistry XPT System Software Test Definition (TDef) Version 1.0 disks (SMN: 11127343 and 11222123). This issue affects only the ADVIA Chemistry Hemoglobin A1c_3 Automated Pretreatment (A1c_3) results when HbA1c values are reported in International Federation of Clinical Chemistry (IFCC) equivalent units (HbA1cR). The ratio equation used to calculate ADVIA Chemistry Hemoglobin A1c_3 Automated Pretreatment (A1c_3) results in IFCC units contains an error. The error results in falsely depressed HbA1cR results.
- Tosoh Bioscience IncTerminatedUnknown
HbAE is known to interfere with the HbA1c assay on the current version of software, Ver. 5.23. Customers should exercise caution when reviewing chromatograms and ensure that Flag 43 is enabled on their device to avoid reporting invalid test results in the presence of HbAE.
- Siemens Healthcare Diagnostics, Inc.TerminatedUnknown
If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer is configured to run Code 39 with check digit and an operator erroneously scans a Code 39 barcode without a check digit, the scanner incorrectly sends the code and the last character is removed.
- Tosoh Bioscience IncTerminatedUnknown
Analyzers were distributed with software which lacks a 510K