Recalls for Dialyzer, High Permeability With Or Without Sealed Dialysate System
216 on file · showing the 200 most recent
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedJul 8, 2026
Potential for Online Port Errors due to connection issues when connecting the SafeLine of the 5008X CAREset to the substitution fluid port of the 5008X CAREsystem.
- VANTIVE US HEALTHCARE LLCOpen, ClassifiedFeb 2, 2026
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
- VANTIVE US HEALTHCARE LLCOpen, ClassifiedFeb 2, 2026
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
- VANTIVE US HEALTHCARE LLCOpen, ClassifiedFeb 2, 2026
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
- VANTIVE US HEALTHCARE LLCOpen, ClassifiedFeb 2, 2026
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
- VANTIVE US HEALTHCARE LLCOpen, ClassifiedFeb 2, 2026
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
- VANTIVE US HEALTHCARE LLCOpen, ClassifiedFeb 2, 2026
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
- VANTIVE US HEALTHCARE LLCOpen, ClassifiedFeb 2, 2026
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedDec 17, 2025
Several software anomalies with the potential to impact patient treatment, of which one includes the possibility to result in treatment stop if certain conditions are fulfilled.
- Nipro Medical CorporationOpen, ClassifiedDec 11, 2025
Screw may loosen or fallout of Hemodialysis system
- VANTIVE US HEALTHCARE LLCOpen, ClassifiedAug 26, 2025
Low deaeration chamber level at post priming tests may lead to PrisMax System alarm T2309: Air Detected in Prime resulting in multiple additional priming cycles
- VANTIVE US HEALTHCARE LLCOpen, ClassifiedAug 26, 2025
Unexpected PrisMax System alarm T0830: Blood Leak Detected, when no blood leak is present, or difficulty normalizing the Blood Leak Detector (BLD) leading to PrisMax System alarms T1313: BLD Normalize Failed, T0853: Normalization Failed, or T1205: BLD Self-Test Failure
- VANTIVE US HEALTHCARE LLCOpen, ClassifiedAug 26, 2025
Thermax System may be unable to detect the presence of a bag on the Thermax Blood Warmer Unit leading to difficulty setting up therapy, or PrisMax System alarm T2284: Thermax Disposable Not Inserted
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedJun 18, 2025
Potential of blood leak occurring between the lower port of the venous chamber and the tubing include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.
- Nuwellis IncCompletedJan 16, 2025
The AquaFlexFlow UF 500 Plus extracorporeal blood circuit used with the Aquadex SmartFlow and FlexFlow Systems may indicate "Ultrafiltrate Weight Mismatch" or Excessive Weight Mismatch Alarms while in use. If not addressed, this failure could result in excess fluid removal from a patient leading to Acute Volume Depletion. This failure is especially serious when the Aquadex System is being used on pediatric patients.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedJan 3, 2025
Potential for internal blood leaks due to cracked polyurethane
- Baxter Healthcare CorporationOpen, ClassifiedNov 13, 2024
Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.
- Outset Medical, Inc.Open, ClassifiedOct 3, 2024
Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if blood loss, less-than 300ml, occurs, the patient may experience hypotension, fatigue, cramps, dizziness, headaches, nausea, and shortness of breath.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedSep 11, 2024
Machines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005) will not record Blood Pressure measurements and does not alarm if the cuff tubing is kinked close to the cuff.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedJul 11, 2024
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedJul 11, 2024
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedJul 11, 2024
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedJul 11, 2024
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedJul 11, 2024
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedJul 11, 2024
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedJul 11, 2024
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedJul 11, 2024
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- Outset Medical, Inc.Open, ClassifiedJun 14, 2024
A cart with prefiltration drawer, an optional hemodialysis system accessory, has a cord with a female connector that has internal contact pins that may become damaged that could create localized heat, sparks, smoke, melting or charring (thermal event) of the female connector or the associated connector power entry module, which may lead to transient delay in treatment or blistering to the user.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedMay 17, 2024
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedMay 17, 2024
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedMay 17, 2024
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedMay 17, 2024
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedMay 17, 2024
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- Outset Medical, Inc.Open, ClassifiedApr 12, 2024
Hemodialysis System, pre and post dialyzer peroxide-cured silicone tubing, with less than 336 hours usage, will be replaced because tubing may contain non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs), that have not been flushed out over time with usage. Exposure to PCBAs could cause: skin conditions, liver damage, impaired reproduction, development of certain kinds of cancer in humans.
- NxStage MDS CorporationOpen, ClassifiedFeb 27, 2024
Product is recalled due to reports of blood leaking or spraying from the arterial line on the cartridge.If this failure occurs, a patient may experience uncontrolled blood loss or a healthcare professional could be exposed to infectious patient blood.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedOct 6, 2023
Potential PCBA leaching from tubing of hemodialysis machines.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedOct 6, 2023
Potential PCBA leaching from tubing of hemodialysis machines.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedOct 6, 2023
Potential PCBA leaching from tubing of hemodialysis machines.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedOct 6, 2023
Potential PCBA leaching from tubing of hemodialysis machines.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedOct 6, 2023
Potential PCBA leaching from tubing of hemodialysis machines.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedOct 6, 2023
Potential PCBA leaching from tubing of hemodialysis machines.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedOct 6, 2023
Potential PCBA leaching from tubing of hemodialysis machines.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedOct 6, 2023
Potential PCBA leaching from tubing of hemodialysis machines.
- Baxter Healthcare CorporationOpen, ClassifiedDec 22, 2022
Potential for Internal Blood Leaks due to a twisted gasket in the dialyzer
- Baxter Healthcare CorporationOpen, ClassifiedDec 22, 2022
Potential for Internal Blood Leaks due to a twisted gasket in the dialyzer
- Baxter Healthcare CorporationOpen, ClassifiedDec 22, 2022
Reports of Polyurethane Displacement in the dialyzer, and might be identified as a blood leak
- Baxter Healthcare CorporationOpen, ClassifiedDec 22, 2022
Reports of Polyurethane Displacement in the dialyzer, and might be identified as a blood leak
- Baxter Healthcare CorporationOpen, ClassifiedDec 22, 2022
Reports of Polyurethane Displacement in the dialyzer, and might be identified as a blood leak
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedDec 20, 2022
(1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false alarm "Remove USB 2 Device".
- Baxter Healthcare CorporationOpen, ClassifiedOct 12, 2022
Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
- Baxter Healthcare CorporationOpen, ClassifiedOct 12, 2022
Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
- Baxter Healthcare CorporationOpen, ClassifiedOct 12, 2022
Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
- Baxter Healthcare CorporationOpen, ClassifiedOct 12, 2022
Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
- Baxter Healthcare CorporationOpen, ClassifiedOct 12, 2022
Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
- Baxter Healthcare CorporationTerminatedOct 30, 2019
TherMax Blood Warmers may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds.
- Fresenius Medical Care Renal Therapies Group, LLCTerminatedMar 20, 2018
When the recirculation ultrafiltration (UF) Goal is set to a value greater than 200ml in service mode and the user starts a treatment using the SLED (Sustained Low Efficiency Dialysis) program the display will show an invalid message and the UF pump will run at the recirculation UF rate which may be up to 4000ml/hour. The SLED program is required to limit the UF rate to a maximum of 1000 ml/hour.
- Fresenius Medical Care Renal Therapies Group, LLCTerminatedFeb 20, 2018
While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.
- B Braun Medical IncTerminatedMay 3, 2016
Potential leakage of conductivity sensors of the Dialog Dialysis machines.
- B Braun Medical IncTerminatedMay 3, 2016
Potential leakage of conductivity sensors of the Dialog Dialysis machines.
- B Braun Medical IncTerminatedMay 3, 2016
Potential leakage of conductivity sensors of the Dialog Dialysis machines.
- B Braun Medical IncTerminatedMay 3, 2016
Potential leakage of conductivity sensors of the Dialog Dialysis machines.
- B Braun Medical IncTerminatedMay 3, 2016
Potential leakage of conductivity sensors of the Dialog Dialysis machines.
- B Braun Medical IncTerminatedMay 3, 2016
Potential leakage of conductivity sensors of the Dialog Dialysis machines.
- Fresenius Medical Care Renal Therapies Group, LLCTerminatedApr 5, 2016
Added Warning Statement: Fresenius 2008 Series Hemodialysis Machines False Blood Leak Alarm When Dialyzing Patients Treated with Hydroxocobalamin (or any form of Vitamin B-12)
- Medivators, Inc.TerminatedSep 11, 2015
Hemoconcentrators may exhibit low ultrafiltration performance that is below product specification.
- Medtronic Perfusion SystemsTerminatedSep 4, 2015
Hemocor High Performance Hemoconcentrators demonstrate low ultrafiltration performance that is below product specification. These potentially affected units have been distributed as stand-alone devices and in specific lots of Medtronic Perfusion Tubing packs.
- Fresenius Medical Care Holdings, Inc.TerminatedApr 29, 2015
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
- Fresenius Medical Care Holdings, Inc.TerminatedApr 29, 2015
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
- Fresenius Medical Care Holdings, Inc.TerminatedApr 29, 2015
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
- Fresenius Medical Care Holdings, Inc.TerminatedApr 29, 2015
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
- Baxter Healthcare Corp.CompletedMar 13, 2015
Product may have separating or protruding sponges
- Fresenius Medical Care Holdings, Inc.TerminatedJan 27, 2015
Unqualified power assemblies shipped to customers.
- Gambro Renal Products IncTerminatedDec 1, 2014
Gambro is initiating a field action for Polyflux Revaclear Dialyzer and Polyflux Revaclear Max Dialyzer due to potential tears in the individual packaging that may compromise the sterile barrier.
- Gambro Renal Products IncTerminatedDec 1, 2014
Gambro is initiating a field action for Polyflux Revaclear Dialyzer and Polyflux Revaclear Max Dialyzer due to potential tears in the individual packaging that may compromise the sterile barrier.
- Gambro Renal Products, IncorporatedTerminatedAug 13, 2014
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
- Gambro Renal Products, IncorporatedTerminatedAug 13, 2014
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
- Gambro Renal Products, IncorporatedTerminatedAug 13, 2014
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
- Gambro Renal Products, IncorporatedTerminatedAug 13, 2014
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
- Gambro Renal Products, IncorporatedTerminatedAug 13, 2014
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
- Gambro Renal Products, IncorporatedTerminatedAug 13, 2014
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
- Gambro Renal Products, IncorporatedTerminatedAug 13, 2014
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
- Gambro Renal Products, IncorporatedTerminatedAug 13, 2014
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
- Gambro Renal Products, IncorporatedTerminatedAug 13, 2014
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
- Gambro Renal Products, IncorporatedTerminatedAug 13, 2014
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
- Gambro Renal Products, IncorporatedTerminatedAug 13, 2014
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
- Baxter Healthcare Corp.TerminatedJul 17, 2014
Inadequate iodine and packaging related defects.
- Baxter Healthcare Corp.TerminatedJun 30, 2014
the white printed paper layer on the pouch packaging may not be fully adhered to the pouch foil layer. The upper paper layer may serve as a protective barrier and help support overall packaging.
- Gambro Renal Products, IncorporatedTerminatedMay 21, 2014
The firm received several complaints for internal blood leaks involving various lot numbers produced in late 2013.
- Gambro Renal Products, IncorporatedTerminatedMay 21, 2014
The firm received several complaints for internal blood leaks involving various lot numbers produced in late 2013.
- Fresenius Medical Care Holdings, Inc.TerminatedMay 13, 2014
2008-Series Hemodialysis Machines with Revision K Actuator Test Boards may fail to detect a disconnected Ultrafiltration (UF) Pump
- Fresenius Medical Care Holdings, Inc.TerminatedMay 13, 2014
2008-Series Hemodialysis Machines with Revision K Actuator Test Boards may fail to detect a disconnected Ultrafiltration (UF) Pump
- Fresenius Medical Care Holdings, Inc.TerminatedFeb 26, 2014
Dialysate Port Leak During Priming of the Dialyzers
- Fresenius Medical Care Holdings, Inc.TerminatedFeb 26, 2014
Dialysate Port Leak During Priming of the Dialyzers
- Fresenius Medical Care Holdings, Inc.TerminatedFeb 26, 2014
Dialysate Port Leak During Priming of the Dialyzers
- Fresenius Medical Care Holdings, Inc.TerminatedFeb 26, 2014
Dialysate Port Leak During Priming of the Dialyzers
- Fresenius Medical Care Holdings, Inc.TerminatedFeb 26, 2014
Dialysate Port Leak During Priming of the Dialyzers
- Fresenius Medical Care Holdings, Inc.TerminatedFeb 26, 2014
Dialysate Port Leak During Priming of the Dialyzers
- Fresenius Medical Care Holdings, Inc.TerminatedFeb 26, 2014
Dialysate Port Leak During Priming of the Dialyzers
- Fresenius Medical Care Holdings, Inc.TerminatedFeb 26, 2014
Dialysate Port Leak During Priming of the Dialyzers
- Fresenius Medical Care Holdings, Inc.TerminatedFeb 26, 2014
Dialysate Port Leak During Priming of the Dialyzers
- Fresenius Medical Care Holdings, Inc.TerminatedFeb 3, 2014
Bicarbonate Jug Adapter distributed prior to 510(k) approval
- Fresenius Medical Care Holdings, Inc.TerminatedJan 6, 2014
2008 Series Hemodialysis Machines complaints of saline bags inappropriately filling during recirculation and setup
- Fresenius Medical Care Holdings, Inc. dba Renal SolutionsTerminatedAug 26, 2013
Product is being recalled because the combination of the current 50 mg vial, the 2008 MeDS pump, and the 2008 MeDS pump tubing Set for CombiSet Bloodlines can lead to a situation where Venofer(R) delivery does not meet dosage accuracy requirements. This may lead to a slight over or under delivery of Venofer(R) that does not pose an acceptable risk to the patient. Please note that this recall
- Fresenius Medical Care Holdings, Inc.TerminatedJul 11, 2013
Incorrect rubber nitrile O-rings distributed instead of the correct material EPDM. (Ethylene Propylene Dine Monomer (M-class) rubber)
- Baxter Healthcare Corp.TerminatedJun 21, 2013
leaking pouches
- Asahi Kasei Medical Co., Ltd., Okatomi PlantTerminatedSep 20, 2012
Asahi became aware of a negative trend for rate of the adverse events that may be associated with a manufacturing change to certain lot numbers of the products, though there is no sufficient information to date to conclusively link these adverse events to the products.
- Asahi Kasei Medical Co., Ltd., Okatomi PlantTerminatedSep 20, 2012
Asahi became aware of a negative trend for rate of the adverse events that may be associated with a manufacturing change to certain lot numbers of the products, though there is no sufficient information to date to conclusively link these adverse events to the products.
- Asahi Kasei Medical Co., Ltd., Okatomi PlantTerminatedSep 20, 2012
Asahi became aware of a negative trend for rate of the adverse events that may be associated with a manufacturing change to certain lot numbers of the products, though there is no sufficient information to date to conclusively link these adverse events to the products.
- Asahi Kasei Medical Co., Ltd., Okatomi PlantTerminatedSep 20, 2012
Asahi became aware of a negative trend for rate of the adverse events that may be associated with a manufacturing change to certain lot numbers of the products, though there is no sufficient information to date to conclusively link these adverse events to the products.
- Asahi Kasei Medical Co., Ltd., Okatomi PlantTerminatedSep 20, 2012
Asahi became aware of a negative trend for rate of the adverse events that may be associated with a manufacturing change to certain lot numbers of the products, though there is no sufficient information to date to conclusively link these adverse events to the products.
- Asahi Kasei Medical Co., Ltd., Okatomi PlantTerminatedSep 20, 2012
Asahi became aware of a negative trend for rate of the adverse events that may be associated with a manufacturing change to certain lot numbers of the products, though there is no sufficient information to date to conclusively link these adverse events to the products.
- Asahi Kasei Medical Co., Ltd., Okatomi PlantTerminatedSep 20, 2012
Asahi became aware of a negative trend for rate of the adverse events that may be associated with a manufacturing change to certain lot numbers of the products, though there is no sufficient information to date to conclusively link these adverse events to the products.
- Fresenius Medical Care Holdings, Inc.TerminatedMar 30, 2012
Risk of internal blood leaks affecting the hemodialyzers.
- Fresenius Medical Care Holdings, Inc.TerminatedJan 20, 2012
Optiflux F250NRe Hemodialyzers have increased risk of an internal blood leak
- Baxter Healthcare Renal DivTerminatedNov 7, 2011
The Tina System was released to the customer with OOS values for the Pressure Displays Accuracy and Leakage Rate Test.
- Gambro Renal Products, IncorporatedTerminatedApr 6, 2011
Prismaflex touch screen may become unresponsive during Continuous Renal Replacement Therapy procedure.
- Fresenius Medical Care Holdings, Inc.TerminatedMar 22, 2011
There is an increased risk of an internal blood leak.
- Fresenius Medical Care North AmericaTerminatedAug 15, 2010
Power cords may fail if the prongs crack or fail at the plug, and may become a fire/burn/shock hazard or delay/interrupt therapy.
- Fresenius Medical Care North AmericaTerminatedJul 1, 2010
Hemodialyzer may leak at the header/end cap
- Edwards Lifesciences, LlcTerminatedJan 27, 2010
There have been reports of clinically significant fluid imbalance when a certain level of fluid imbalance is detected the Aquarius will trigger an alarm, however, users are able to override this alarm and continue therapy. It is possible to remove too much fluid from or replace too much fluid to the patient. In extreme cases, this may result in serious injuries or death.
- Baxter Healthcare Renal DivTerminatedNov 12, 2009
When the main power is lost temporarily, a software anomaly prevents the instrument from automatically changing the ultrafiltration rate (UFR) to the minimum value as described in the Operator's manual.
- Gambro Renal Products, Inc.TerminatedMar 25, 2008
Kinked tubing on hemodialysis device may cause hemolysis.
- Fresenius Medical Care North AmericaTerminatedFeb 1, 2008
Leaking: the Dialyzer may leak at the header resulting in small amounts of blood loss.
- NxStage Medical, Inc.TerminatedSep 20, 2007
Potential for dialysate leaks from the cartridge during treatment.
- Gambro Renal Products, Inc.TerminatedMar 31, 2007
Under certain conditions, * an excessive amount of anticoagulant may be infused into the patient. (* interruption of a self-test by an alarm and obstruction of the access lines)
- Gambro Renal Products, Inc.TerminatedNov 21, 2006
Defective Heat Exchanger/membrane; this may result in perforation of the membrane thus allowing the presence of infectious organisms and/or pathogens to pass from patient to patient.
- Baxter Healthcare Renal DivTerminatedNov 14, 2006
The Meridian pump is less likely to detect small pressure changes, which may indicate the presence of a post-pump tubing kink, when using pre-pump arterial pressure monitoring.
- Asahi Medical Co., LtdTerminatedNov 14, 2006
There have been reports of difficulty separating the Gambro Cartridge Blood Set from Asahi Reuse/Multiple Use Dialyzers at the end of treatment, possibly damaging the dialyzer nozzle.
- Asahi Medical Co., LtdTerminatedNov 14, 2006
There have been reports of difficulty separating the Gambro Cartridge Blood Set from Asahi Reuse/Multiple Use Dialyzers at the end of treatment, possibly damaging the dialyzer nozzle.
- Gambro Renal Products, Inc.TerminatedOct 12, 2006
A potential risk of infusing air or infusion fluid to the patient during dialysis. Also, a possibility of an incorrect scale reading of the amount of patient fluid removal.
- Baxter Healthcare Renal DivTerminatedJul 25, 2006
The UF Removal Regulators were assembled incorrectly. The mis-assembly results in a failure to successfully calibrate the instrument and begin patient therapy.
- Aksys, Ltd.TerminatedJul 13, 2006
The hemodialysis device was being marketed with a modified treatment length that exceeds the treatment length for which the device was originally designed.
- Gambro Renal Products, Inc.TerminatedApr 5, 2006
After dialysis machine is disconnected from the patient, blood leakage may occur while unloading the tubing sets from the machine.
- Gambro Renal Products, Inc.TerminatedApr 5, 2006
After dialysis machine is disconnected from the patient, blood leakage may occur while unloading the tubing sets from the machine.
- Baxter Healthcare Renal DivTerminatedMar 29, 2006
Advisory to alert users to assure that the hemofilter of the blood line is in the horizontal position to avoid kinking of the bloodline tubing to avoid hemolysis, and to show the compatible blood tubing sets to be used with the BM11/BM11A Blood Monitor Pump and BM25 System ultrafiltration hemodialysis system.
- Baxter Healthcare Renal DivTerminatedMar 29, 2006
Advisory to alert users to assure that the hemofilter of the blood line is in the horizontal position to avoid kinking of the bloodline tubing to avoid hemolysis, and to show the compatible blood tubing sets to be used with the BM11/BM11A Blood Monitor Pump and BM25 System ultrafiltration hemodialysis system.
- Baxter Healthcare Renal DivTerminatedJan 6, 2006
Potential for fluid or air to be passed through the venous line clamp if the tubing is not centered in the clamp and extends beyond the edge of the clamp's pinch zone. This could result in an air emboli condition.
- Fresenius Medical Care North AmericaTerminatedDec 31, 2005
Mislabeled: Inner product mislabeled as F180NR (Catalog Number 0500318N) instead of F160NR
- Fresenius Medical Care North AmericaTerminatedDec 31, 2005
Mislabeled: Inner product mislabeled as 180NR (Catalog Number 0500318N) instead of 160NR
- Baxter Healthcare Renal DivTerminatedDec 14, 2005
Inadequate instructions for routing single tubing lines through the double-tubing clips on the front of the instrument to eliminate the risk of kinks in the tubing, which have been associated with a death and a serious injury.
- Baxter Healthcare Renal DivTerminatedNov 16, 2005
A software anomaly that occurs when the Reset Totals option is selected during CVVHD therapy with a diaylsate flow rate of less than 600 mL/hour, or CVVH therapy with a predilution flow rate of less than 600 mL/hour, causes balance alarms without an obvious cause.
- Gambro Renal ProductsTerminatedOct 1, 2005
Critically ill patients undergoing continuous renal replacement therapy may suffer excessive fluid loss.
- Baxter Healthcare Renal DivTerminatedAug 24, 2005
Operator's Manuals addenda are being issued to provide correct information concerning the correct syringe size to use for heparin infusion, the use of chemical disinfectants in the System 1000, TMP Advise, and Bicarbonate Value calculations.
- Baxter Healthcare Renal DivTerminatedAug 18, 2005
In the event therapy is terminated, as part of routine instrument programming or therapy maintenance, the Accura instrument may experience difficulty in performing a successful self-test (failure) upon an immediate attempt to restart therapy
- Baxter Healthcare Renal DivTerminatedJun 30, 2005
The saline bags may inappropriately empty/fill during the recirculation mode of the hemodialysis system.
- Asahi Medical Co LtdTerminatedMay 3, 2005
Exposure to freezing temperatures may damage the hollow fiber dialyzers and result in blood leaks during dialysis.
- Asahi Medical Co LtdTerminatedMay 3, 2005
Exposure to freezing temperatures may damage the hollow fiber dialyzers and result in blood leaks during dialysis.
- Asahi Medical Co LtdTerminatedMay 3, 2005
Exposure to freezing temperatures may damage the hollow fiber dialyzers and result in blood leaks during dialysis.
- Asahi Medical Co LtdTerminatedMay 3, 2005
Exposure to freezing temperatures may damage the hollow fiber dialyzers and result in blood leaks during dialysis.
- Asahi Medical Co LtdTerminatedMay 3, 2005
Exposure to freezing temperatures may damage the hollow fiber dialyzers and result in blood leaks during dialysis.
- Baxter Healthcare Renal DivTerminatedDec 8, 2004
The instrument power cord may become disconnected from the instrument and pose a fire or electrical shock hazard.
- Baxter Healthcare Renal DivTerminatedAug 11, 2004
The Accura machine display screen may go blank during patient treatment. There is a potential risk to the patient in that the operator may try to manipulate the controls in response to a blank screen, which could affect the current patient's treatment
- Baxter Healthcare Renal DivTerminatedJul 20, 2004
These devices have the potential to overheat in the absence of a Ground Fault Circuit Interrupter (GFCI). This heater system failure could result in a fire.
- Asahi Medical Co LtdTerminatedJul 20, 2004
Customer reprocessing methods for the reuseable hollow fiber dialyzers may damage the hollow fibers and result in blood leaks during dialysis.
- Baxter Healthcare Renal DivTerminatedJul 20, 2004
Microbubbles of air in the extracorporeal blood circuit during hemodialysis therapy on the Meridian.
- NxStage Medical, Inc.TerminatedJul 20, 2004
Pinhole leaks on the replacement (RF) side FMP bag in certain lots of its disposable cartridge balance chambers resulting in loss of dialysate or ultrafiltrate
- NxStage Medical, Inc.TerminatedJul 20, 2004
Pinhole leaks on the replacement (RF) side FMP bag in certain lots of its disposable cartridge balance chambers resulting in loss of dialysate or ultrafiltrate
- NxStage Medical, Inc.TerminatedJul 20, 2004
Pinhole leaks on the replacement (RF) side FMP bag in certain lots of its disposable cartridge balance chambers resulting in loss of dialysate or ultrafiltrate
- Baxter Healthcare Corp.TerminatedDec 4, 2003
Use of the dialyzers may cause iritis (red-eye) patient reactions.
- Baxter Healthcare Corp.TerminatedMar 5, 2003
Meridian hemodialysis instrument may reset when in shutdown mode causing the system to reset with potential for patient increased bioburden exposure.
- Baxter Healthcare CorporationOpen, ClassifiedUnknown
If the operator initiates therapy with a saved prescription profile and makes a change to the prescription after a disposable filter change using the Same Patient button, the system may suggest values from the original prescription profile, rather than the current prescription.
- Baxter Healthcare CorporationOpen, ClassifiedUnknown
Communication error alarms may result in interruption of therapy, delay in therapy, or blood loss due to non-restitution of blood in the extracorporeal circuit.
- Baxter Healthcare CorporationOpen, ClassifiedUnknown
Communication error alarms may result in interruption of therapy, delay in therapy, or blood loss due to non-restitution of blood in the extracorporeal circuit.
- Baxter Healthcare CorporationOpen, ClassifiedUnknown
Communication error alarms may result in interruption of therapy, delay in therapy, or blood loss due to non-restitution of blood in the extracorporeal circuit.
- Baxter Healthcare CorporationOpen, ClassifiedUnknown
Communication error alarms may result in interruption of therapy, delay in therapy, or blood loss due to non-restitution of blood in the extracorporeal circuit.
- Baxter Healthcare CorporationOpen, ClassifiedUnknown
Communication error alarms may result in interruption of therapy, delay in therapy, or blood loss due to non-restitution of blood in the extracorporeal circuit.
- NxStage Medical, Inc.TerminatedUnknown
Ultrafiltration (UF) Volume software error inaccurate fluid removal
- NxStage Medical, Inc.TerminatedUnknown
Ultrafiltration (UF) Volume software error inaccurate fluid removal
- NxStage Medical, Inc.TerminatedUnknown
Ultrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF volume has been removed
- NxStage Medical, Inc.TerminatedUnknown
Ultrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF volume has been removed
- NxStage Medical, Inc.TerminatedUnknown
Ultrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF volume has been removed
- Nikkiso Ltd - Shizuoka PlantOpen, ClassifiedUnknown
Device lacks premarket approval.
- Fresenius Medical Care Renal Therapies Group, LLCTerminatedUnknown
Potential for external blood leaks from the dialyzer header
- Outset Medical, Inc.Open, ClassifiedUnknown
Due to a component in the hemodialysis console there is the possibility of heat-related damage that may occur with the device.
- B Braun Medical IncTerminatedUnknown
Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.
- B Braun Medical IncTerminatedUnknown
Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.
- Baxter Healthcare CorporationOpen, ClassifiedUnknown
This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.
- Baxter Healthcare CorporationOpen, ClassifiedUnknown
This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.
- Baxter Healthcare CorporationOpen, ClassifiedUnknown
This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.
- Fresenius Medical Care Renal Therapies Group, LLCTerminatedUnknown
When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine
- Fresenius Medical Care Renal Therapies Group, LLCTerminatedUnknown
When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine
- Fresenius Medical Care Renal Therapies Group, LLCTerminatedUnknown
When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine
- Fresenius Medical Care Renal Therapies Group, LLCTerminatedUnknown
When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine
- Baxter Healthcare CorporationTerminatedUnknown
Baxter Healthcare Corporation is issuing a voluntary product recall for several lots of Phoenix Dialysis Machine Preventive Maintenance Kits due to incorrect silicone straight connectors components packaged in these kits. During Preventive Maintenance, the two silicone connectors must be installed in the Phoenix mass-balance section of hydraulic circuit only if the Phoenix machine has already been modified with "CONN. IMPROVING RETR. KIT" (Code SP00721). Installation of these connectors in machines modified with SP00721 may result in fluid leaks during dialysis treatment. These Preventive Maintenance Kits include the FILTER sub-group, which contains the silicone straight connectors that are larger than expected.
- Baxter Healthcare CorporationTerminatedUnknown
Baxter Healthcare Corporation is issuing a voluntary product recall for several lots of Phoenix Dialysis Machine Preventive Maintenance Kits due to incorrect silicone straight connectors components packaged in these kits. During Preventive Maintenance, the two silicone connectors must be installed in the Phoenix mass-balance section of hydraulic circuit only if the Phoenix machine has already been modified with "CONN. IMPROVING RETR. KIT" (Code SP00721). Installation of these connectors in machines modified with SP00721 may result in fluid leaks during dialysis treatment. These Preventive Maintenance Kits include the FILTER sub-group, which contains the silicone straight connectors that are larger than expected.
- Baxter Healthcare CorporationTerminatedUnknown
Variability in the performance of the tubing in the ARPS (Automatic Repositioning System) Pump Assembly which may lead to alarm situations during system self-test and therapy. This tubing is normally replaced during annual preventative maintenance. Alarm Situations: During System Self-Test: Alarm code B1215 (ARPS Self-Test failure); During Therapy: B1273 (ARPS Leak), B1215 (ARPS Pressure self-test), B1116 (Return Pressure self-test Failure), B1539 (Access Pod Reposition Failure), B1541 (Effluent Pod Reposition Failure), B1540 (Effluent Filter Pod Reposition Failure). If an alarm occurs, it may lead to delay or interruption of therapy. In the event that therapy is terminated without returning blood to the patient, blood loss may occur.
- Baxter Healthcare CorporationTerminatedUnknown
Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.
- Baxter Healthcare CorporationTerminatedUnknown
Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.
- Baxter Healthcare CorporationTerminatedUnknown
Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.
- Baxter Healthcare CorporationTerminatedUnknown
Firm has received reports of device operators failing to adhere to instructions for use pertaining to the safe unloading of disposable sets from the Prismaflex Control Unit. Additionally, for software versions 5.10 and 6.10, the programmed syringe size for the syringe pump may revert to safe default values unintentionally.
- Baxter Healthcare CorporationTerminatedUnknown
Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.
- Fresenius Medical Care Holdings, Inc.TerminatedUnknown
Dialyzers were distributed after being exposed to cold temperatures which exceeded the low-temperature range in the firm's standard operating procedure.
- Baxter Healthcare CorporationTerminatedUnknown
There is the potential presence of particular matter in the header caps of ten lots the Revaclear 300 Dialyzer.
- Fresenius Medical Care Holdings, Inc.TerminatedUnknown
The firm is replacing pressure transducers with an improved design due to an increase in customer complaints regarding lower long-term reliability of the part.
- Fresenius Medical Care Renal Therapies Group, LLCTerminatedUnknown
Power Logic Board Fail Message on the 2008T BlueStar Machine with software version 2.72.
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedUnknown
A "Remove USB Device 2" false alarm may be displayed when no USB device or a non-powered USB device is connected to the USB port on the rear of the machine
- Fresenius Medical Care Holdings, Inc.Open, ClassifiedUnknown
A "Remove USB Device 2" false alarm may be displayed when no USB device or a non-powered USB device is connected to the USB port on the rear of the machine
- Baxter Healthcare CorporationTerminatedUnknown
Baxter is communicating safety information from the Prismaflex Operator's Manual regarding the proper use of connectors with the Prismaflex system.
- Baxter Healthcare CorporationTerminatedUnknown
Baxter is communicating safety information from the Prismaflex Operator's Manual regarding the proper use of connectors with the Prismaflex system.
- Baxter Healthcare CorporationTerminatedUnknown
Baxter is communicating safety information from the Prismaflex Operator's Manual regarding the proper use of connectors with the Prismaflex system.