Hemoglobin A2 Quantitation
Hematology · 21 CFR 864.7400
Recent MDR events
0 adverse event reports on file
No MDR events on file for this product code.
Event type breakdown
No data available.
Most-reported manufacturers
No data available.
Recalls
5 recall events for this product code
- Bio-Rad Laboratories, Inc.TerminatedMay 2, 2014
Variant II beta-thalassemia Short Program Model Numbers 270-2103 (250 tests) and 270-2154 (500 tests), Bio-Rad Laboratories, Inc. in vitro diagnostic. The Var
There is an issue with the buffers which impacts the quantification of the Hemoglobin F peak. Calibration and Quality control are likely to fail if the affected lots are used.
Z-1537-2014 · View FDA record ↗ - Tosoh Bioscience IncTerminatedApr 10, 2013
G7 HPLC.
The Beta-thalessemia Mode of the G7 Analyzer-Beta Thalassemia Mode is an automated High Performance Liquid Chromatography (HPLC) system that separates and reports HbF and HbA2 quantitative percentages in whole blood. The firm received complaints that the A2 peak could not be detected with the G7 Beta-thalessemia columns.
Z-1075-2013 · View FDA record ↗ - Tosoh Bioscience IncTerminatedAug 17, 2011
Hemoglobin F & A2 Calibrator,
The value after reconstitution appeared to be lower than the assigned value on the label. The measured value for Lot BCAL05R was 5.3%, whereas the value on the label is 5.7%. The criteria of the value is +/- 0.3%.
Z-3030-2011 · View FDA record ↗ - Bio-Rad Laboratories IncTerminatedSep 11, 2008
VARIANT -thalassemia Short Program Hemoglobin Testing System, Model Number 270-0019 (250 tests), with -thalassemia ROM card lot PA 62600, PIN 2700104
Some product ROM cards have an incorrect program loaded and are unable to be updated.
Z-1588-2008 · View FDA record ↗ - Bio-Rad Laboratories IncTerminatedSep 11, 2008
VARIANT -thalassemia Short Program Hemoglobin Testing System, Model Number 270-0002 (500 tests), with -thalassemia ROM card lot PA 62600, PIN 2700104
Some product ROM cards have an incorrect program loaded and are unable to be updated.
Z-1589-2008 · View FDA record ↗
Registered firms
9 firms currently registered for this product code
510(k) clearances
13 clearances for this product code
- Tosoh Medics, Inc.May 14, 2002
K020489 · G7 AUTOMATED HPLC ANALYZER, BETA-THALASSEMIA MODE
- Bio-RadJun 10, 1999
K991127 · VARIANT II B-THALASSEMIA
- Bio-RadMar 11, 1993
K924122 · VARIANT BETA-THALASSEMIA SHORT PROGRAM
- Drew Scientific , Ltd.Nov 16, 1992
K924496 · GLYCOMAT GREEN HBA2 ESTIMATION REAGENT KIT
- Helena LaboratoriesAug 25, 1992
K923317 · COLUMNMATE BETA THAL HBA2 CONTROL-ABNORMAL
- Helena LaboratoriesAug 25, 1992
K923318 · COLUMNMATE BETA-THAL HBA2 CONTROL - NORMAL
- Helena LaboratoriesDec 17, 1991
K913384 · COLUMNMATE(TM) BETA-THAL CAT. NO. 4515
- Bio-RadJan 18, 1990
K896168 · HEMOGLOBIN A2 MICRO COLUMN TEST