Device, Monitoring, Intracranial Pressure
Neurology · 21 CFR 882.1620
Recent MDR events
6,841 adverse event reports on file
- EDM Lumbar Drainage KitJul 30, 2026MalfunctionMedtronic Mexico S. De R.l. De Cv9617601-2026-04764
- EDM Lumbar Drainage KitJul 28, 2026MalfunctionMedtronic Mexico S. De R.l. De Cv9617601-2026-04656
- Pressio ICP monitoring kit, ventricular tunnelingJul 27, 2026OtherSophysa3001587388-2026-00071
- NEUROMONITOR BASIC KITJul 27, 2026MalfunctionIntegra Lifesciences Mansfield3013886523-2026-00186
- PRESSIO® ICP AND ICT MONITORING KITJul 21, 2026MalfunctionSophysa3001587388-2026-00067
- EDM LUMBAR DRAINAGE KITJul 21, 2026InjuryMedtronic Mexico S. De R.l. De Cv9617601-2026-04479
- NEUROVENT-PJul 21, 2026InjuryRaumedic Ag3006942548-2026-00007
- PRESSIO ICP MONITORING KIT, VENTRICULAR TUNNELINGJul 17, 2026MalfunctionSophysa3001587388-2026-00065
- NEUROMONITOR BASIC KITJul 17, 2026InjuryIntegra Lifesciences Mansfield3013886523-2026-00171
- CERELINK ICP MONITOR, POOLJul 17, 2026MalfunctionIntegra Lifesciences Mansfield3014334038-2026-00058
Event type breakdown
Recalls
52 recall events for this product code
- SophysaOpen, ClassifiedJan 8, 2026
The Pressio 2 ICP Monitoring System, Model Number PSO-4000; Intracranial Pressure Monitor
Customer complaints of Pressio monitor rebooting.
Z-1029-2026 · View FDA record ↗ - Integra LifeSciences Corp. (NeuroSciences)Open, ClassifiedJan 8, 2026
Codman Microsensor Basic Kit. Catalog Number: 626631US.
Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits.
Z-1036-2026 · View FDA record ↗ - Integra LifeSciences Corp. (NeuroSciences)Open, ClassifiedJan 8, 2026
Codman CereLink ICP Sensor Basic Kit. Catalog Number: 826850.
Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits.
Z-1037-2026 · View FDA record ↗ - Medtronic NeurosurgeryOpen, ClassifiedDec 19, 2024
Becker External Drainage and Monitoring System:
Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).
Z-0661-2025 · View FDA record ↗ - Medtronic NeurosurgeryOpen, ClassifiedDec 19, 2024
Exacta External Drainage and Monitoring System:
Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).
Z-0662-2025 · View FDA record ↗ - Smiths Medical ASD, Inc.Open, ClassifiedOct 23, 2024
medex TranStar Single OR Set 1/EA, Product Code REF MX20846R2
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0182-2025 · View FDA record ↗ - Smiths Medical ASD, Inc.Open, ClassifiedOct 23, 2024
medex TranStar Single ICP Set 1/EA, Product Code REF MX20897
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Z-0183-2025 · View FDA record ↗ - Sophysa S.A.Open, ClassifiedMay 15, 2024
The Sophysa Pressio Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure with tunneling (PSO
An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.
Z-1825-2024 · View FDA record ↗ - Sophysa S.A.Open, ClassifiedMay 15, 2024
The Sophysa Pressio Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure and temperature wit
An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.
Z-1826-2024 · View FDA record ↗ - Integra LifeSciences Corp.Open, ClassifiedDec 14, 2023
CereLink ICP Extension Cable (Extension Cable), Model No. 826845
Firm is expanding the existing recall RES 90457 for the Cerelink, initiated in June 2022, to include all CereLink ICP Extension Cables. The updated corrective action includes a design change to the extension cable to address the "out of range readings" issue.
Z-0548-2024 · View FDA record ↗
Registered firms
207 firms currently registered for this product code
Sterigenics Us Llc
Los Angeles, CA, US
Isomedix Operations, Inc.
Temecula, CA, US
Reco Medizintechnik Wolfgang Rentsch Ek
Pirna Saxony, DE
Integra Lifesciences Switzerland Sarl
Le Locle Neuchatel, CH
Hipp Medical Ag
Renquishausen Baden-Wurttemberg, DE
Rudolf Storz Gmbh
EMMINGEN Baden-Wurttemberg, DE
Medtronic Neurosurgery
Fort Worth, TX, US
M Dialysis Ab
Stockholm, SE
Ametek Emc
Reynosa Tamaulipas, MX
Blueprint Manufacturing Corporation
Irvine, CA, US
Integra Lifesciences Production Corporation
Mansfield, MA, US
Clearpoint Neuro Inc.
Carlsbad, CA, US
Raumedic - Muenchberg
Helmbrechts Bavaria, DE
Braincare Desenvolvimento E Inovacao Tecnologica S.a.
Sao Carlos Sao Paulo, BR
Sophysa
BESANCON Doubs, FR
510(k) clearances
105 clearances for this product code
- SophysaApr 10, 2026
K252241 · Pressio 3 Multi-parameter Neuromonitoring System
- Raumedic AGAug 29, 2025
K250285 · NEUROVENT-P (092946-003); NEUROVENT (092956-003); NEUROVENT-IFD-R (095317-003); NEUROVENT-IFD-S (091678-003); NEUROVENT-P-TEMP (094268-003); NEUROVENT-TEMP (094278-003); NEUROVENT-TEMP-IFD-R (095327-003); NEUROVENT-TEMP-IFD-S (094288-003); NEUROVENT-PTO (095008-003); NEUROVENT-PTO 2L (095108-003); NEUROVENT-PX (091580-003); BOLT-DRILL KIT CH5 (091888-003); BOLT-DRILL KIT CH9 (091898-003); BOLT-DRILL KIT PTO (092380-003); BOLT-DRILL KIT VP 16 (092969-003); BOLT KIT CH5 (091868-003); BOL
- Braincare Desenvolvimento E Inovação Tecnológica S.A.Jul 18, 2024
K240821 · B4C System - Addition of new sensor model (BcSs-PICNIW-2000)
- Integra Lifesciences Production CorporationFeb 4, 2024
K232890 · CereLink ICP Monitor (826820), CereLink ICP Extension Cable (826845)
- Braincare Desenvolvimento E Inovacao Tecnologica S.A.Oct 8, 2021
K201989 · B4C System
- Integra Lifesciences Production CorporationApr 30, 2021
K210993 · CereLink ICP Monitor
- Medtronic, Inc.May 26, 2020
K201118 · Exacta External Drainage and Monitoring System, Exacta Pole with Laser Level, Exacta System Replacement Drainage Bag, Replacement Laser Level
- Medtronic, Inc.Apr 9, 2020
K200630 · Medtronic External Drainage and Monitoring System, Medtronic External Drainage and Monitoring System, Becker Replacement Drainage Bags