Antisera, Latex Agglutination, Cryptococcus Neoformans
Microbiology · 21 CFR 866.3165
Recent MDR events
2 adverse event reports on file
Event type breakdown
Most-reported manufacturers
Recalls
3 recall events for this product code
- Immuno-Mycologics, IncTerminatedJan 27, 2017
Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003
The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.
Z-1109-2017 · View FDA record ↗ - Immuno-Mycologics, IncTerminatedUnknown
CRYPTOCOCCAL ANTIGEN LATERAL FLOW ASSAY, IMMY, Lateral Flow Assay for the Detection of Cryptococcal Antigen in serum and CSF, REF CR2003, IVD, UDI: (01)008163
The firm found immunoassay products used to detect cryptococcal antigen in serum and cerebrospinal fluid to have reduced specificity (90% now versus 99% before). As a result, a small number of samples with positive test results may be false positives, which may cause some patients to initiate unnecessary anti-fungal therapy. Samples with negative test results are NOT affected.
Z-0545-2022 · View FDA record ↗ - Immuno-Mycologics, IncTerminatedUnknown
Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Model # CR2003
Potential for LFA strips to improperly wick which may cause false negative cryptococcosis test results
Z-2409-2018 · View FDA record ↗
Registered firms
11 firms currently registered for this product code
510(k) clearances
10 clearances for this product code
- Immuno-Mycologics, Inc.Mar 28, 2012
K112422 · CRAG LATERAL FLOW ASSAY (LFA)
- Immuno-Mycologics, Inc.Jul 20, 2011
K102286 · CRAG LATERAL FLOW ASSAY (CRAG LFA)
- Trinity Laboratories, Inc.Feb 17, 1994
K930418 · CRYPTO-LEX SYSTEM
- Alexon Biomedical, Inc.Nov 22, 1993
K933997 · PROSPECT ENTAMOEBA HISTOLYTICA MICROPLATE ASSAY
- Meridian Diagnostics, Inc.Sep 8, 1989
K894845 · (CALAS(TM)) W/ADD'L REAGENT CALAS(TM) PRONASE
- American Scientific ProductsAug 31, 1981
K812188 · CRYPTOCOCCAL ANTIGEN LATEX AGGLUTI
- Immuno-Mycologics, Inc.Mar 20, 1981
K810510 · YA-CRYPTO ANTIBODY TUBE AGGLUTINATION
- Meridian Diagnostics, Inc.Jan 21, 1980
K792693 · CRYPTOCOCCUS NEOFORMANS ANTISERUM, LATEX