Recalls for Light, Ultraviolet, Dermatological
15 on file
- Daavlin Distributing CompanyOpen, ClassifiedJul 15, 2026
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
- Daavlin Distributing CompanyOpen, ClassifiedJul 15, 2026
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
- Daavlin Distributing CompanyOpen, ClassifiedJul 15, 2026
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
- Daavlin Distributing CompanyOpen, ClassifiedJun 2, 2026
Some of the Calibration/Output Certificates had the outputs reversed.
- National Biological CorpTerminatedMar 20, 2017
The wiring may allow the lamps to turn on with the key switch rather than the timer.
- National Biological CorpTerminatedJul 19, 2013
The incorrect manual may have been packaged with this product. UVB-080 devices may have included the manual for a UVA-083 device.
- National Biological CorpTerminatedJul 19, 2013
The incorrect manual may have been packaged with this product. UVB-083 devices may have included the manual for a UVA-080 device.
- National Biological CorpTerminatedAug 28, 2008
Software allows operator to override "low line voltage" error warning and store light intensity value.
- Daavlin Distributing CompanyTerminatedJul 30, 2004
Software errors cause the unit to not deliver the prescribed therapy and do not allow the patient records to be saved.
- Daavlin Distributing CompanyTerminatedJul 30, 2004
Software errors cause the unit to not deliver the prescribed therapy and do not allow the patient records to be saved.
- Daavlin Distributing CompanyTerminatedJul 30, 2004
Software errors cause the unit to not deliver the prescribed therapy and do not allow the patient records to be saved.
- Daavlin Distributing CompanyTerminatedJul 30, 2004
Software errors cause the unit to not deliver the prescribed therapy and do not allow the patient records to be saved.
- Daavlin Distributing CompanyTerminatedJul 30, 2004
Software errors cause the unit to not deliver the prescribed therapy and do not allow the patient records to be saved.
- Daavlin Distributing CompanyOpen, ClassifiedUnknown
Software issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0" inches to "9. Resulting in over or under dosing.
- Daavlin Distributing CompanyOpen, ClassifiedUnknown
Software issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0" inches to "9. Resulting in over or under dosing.