Endoscopic Video Imaging System/component, Gastroenterology-urology
Gastroenterology, Urology · 21 CFR 876.1500
To allow for visualization of body cavities through an endoscope by projecting images to a monitor.
Recent MDR events
61,448 adverse event reports on file
- HD AUTOCLAVABLE CAMERA HEADJul 31, 2026MalfunctionShirakawa Olympus Co., Ltd.3002808148-2026-26371
- EVIS X1 VIDEO SYSTEM CENTERJul 31, 2026MalfunctionShirakawa Olympus Co., Ltd.3002808148-2026-26373
- HD CAMERA HEADJul 31, 2026MalfunctionShirakawa Olympus Co., Ltd.3002808148-2026-26374
- HD AUTOCLAVABLE CAMERA HEADJul 31, 2026MalfunctionShirakawa Olympus Co., Ltd.3002808148-2026-26376
- VIDEO SYSTEM CENTERJul 31, 2026MalfunctionShirakawa Olympus Co., Ltd.3002808148-2026-26378
- EVIS X1 VIDEO SYSTEM CENTERJul 31, 2026MalfunctionShirakawa Olympus Co., Ltd.3002808148-2026-26379
- VISERA ELITE II VIDEO SYSTEM CENTERJul 31, 2026MalfunctionShirakawa Olympus Co., Ltd.3002808148-2026-26381
- HD 3CMOS AUTOCLAVABLE CAMERA HEADJul 31, 2026MalfunctionShirakawa Olympus Co., Ltd.3002808148-2026-26384
- VIDEO SYSTEM CENTERJul 31, 2026MalfunctionShirakawa Olympus Co., Ltd.3002808148-2026-26386
- VISERA ELITE III VIDEO SYSTEM CENTERJul 31, 2026MalfunctionShirakawa Olympus Co., Ltd.3002808148-2026-26387
Event type breakdown
Recalls
8 recall events for this product code
- Olympus Corporation of the AmericasOpen, ClassifiedOct 23, 2024
Olympus Gastrointestinal Videoscope, Model/Catalog Number: GIF-1100
It was discovered during device performance testing that the CCD imaging sensors were programmed with the incorrect color correction data and therefore, specifications are not met.
Z-0181-2025 · View FDA record ↗ - Aizu Olympus Co., Ltd.Open, ClassifiedNov 17, 2023
Colonoscope: for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve).
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Z-0329-2024 · View FDA record ↗ - Aizu Olympus Co., Ltd.Open, ClassifiedNov 17, 2023
Duodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Z-0346-2024 · View FDA record ↗ - Karl Storz EndoscopyTerminatedJul 6, 2016
Image1 SPIES X-LINK Camera Control Unit (CCU), Product Model No. TC301, when used with Image1 SPIES D1 camera head, Product Model No. TH116 & TH115, and DC1 Vid
Potential electrical interference issue when the Image1 SPIES X-LINK and Image1 SPIES D1 camera head are used in conjunction with the DCI Video Mediastinocope and an electro-surgery generator (ESG).
Z-2134-2016 · View FDA record ↗ - Fujifilm Medical Systems U.S.A., Inc.TerminatedSep 22, 2015
Fujinon/Fujifilm Processor EPX 2500 Operation Manual. For use in endoscopic procedures.
The EPX-2500 Operation Manual and sales brochure incorrectly identify that the ED-530XT endoscope is meant to be used with the EPX-2500 Processor. The Operation Manual and sales brochure state that when the ED-530XT endoscope is used with the 2500 processor, a "super image" is displayed when it is actually a "standard image" that is displayed.
Z-2805-2015 · View FDA record ↗ - Fujifilm Medical Systems U.S.A., Inc.TerminatedSep 22, 2015
Fujifilm Processor EPX 2500 High Definition Endoscopy; (Sales Brochure).
The EPX-2500 Operation Manual and sales brochure incorrectly identify that the ED-530XT endoscope is meant to be used with the EPX-2500 Processor. The Operation Manual and sales brochure state that when the ED-530XT endoscope is used with the 2500 processor, a "super image" is displayed when it is actually a "standard image" that is displayed.
Z-2806-2015 · View FDA record ↗ - Olympus Corporation of the AmericasTerminatedUnknown
Olympus JF-140F duodenoscopes (Model NumberJF-140F)
Issuance of validated, new reprocessing procedures. This corrective action is being undertaken as a result of ongoing manufacturer and FDA activities relating to reported patient infections associated with duodenoscopes. The new cleaning procedure requires additional recess flushing and forceps elevator raising/lowering steps during precleaning, manual cleaning and manual disinfection.
Z-2269-2018 · View FDA record ↗ - Olympus Corporation of the AmericasTerminatedUnknown
Olympus JF-160F duodenoscopes (Model NumberJF-160F)
Issuance of validated, new reprocessing procedures. The new cleaning procedure requires additional recess flushing and forceps elevator raising/lowering steps during precleaning, manual cleaning and manual disinfection.
Z-2270-2018 · View FDA record ↗
Registered firms
278 firms currently registered for this product code
Karl Storz Se & Co. Kg
TUTTLINGEN Baden-Wurttemberg, DE
International Medical Lasers, Inc.
Tualatin, OR, US
Olympus Medical Systems Corp.
HACHIOJI-SHI Tokyo, JP
Ambu A/s
Ballerup Hovedstaden, DK
Karl Storz Endoscopy-america, Inc.
El Segundo, CA, US
Plexus Manufacturing Sdn .bhd.
Bayan Lepas Pulau Pinang, MY
Medtronic Neurosurgery
Fort Worth, TX, US
Expeditors International Bv
Amsterdam Noord-Holland, NL
Life Science Logistics, Llc
Brownsburg, IN, US
Ambu Sdn.bhd.
PENANG Pulau Pinang, MY
Probo Medical Ltd
Gloucester Gloucestershire, GB
Shirakawa Olympus Co., Ltd.
NISHISHIRAKAWA-GUN Fukushima, JP
Mdtglobal, A.s.
Bratislava Bratislavsky, SK
Karl Storz Se & Co Kg
TUTTLINGEN Baden-Wurttemberg, DE
Karl Storz Video Endoscopy Estonia Oue
Laagri Harjumaa, EE
Aizu Olympus Co., Ltd.
AIZUWAKAMATSU-SHI Fukushima, JP
Gyrus Acmi, Inc.
Brooklyn Park, MN, US
Arvato Se
Hamburg, DE
Karl Storz Imaging, Inc.
Goleta, CA, US
510(k) clearances
100 clearances for this product code
- Fujifilm CorporationJul 2, 2026
K261880 · FUJIFILM Processor EP-8000;Balloon Controller PB-30;Endoscopy Support Program EW10-VM01
- Olympus Medical Systems CorporationJun 24, 2026
K261260 · HD Camera Head OLYMPUS CH-S200-08-LB
- Proverum LimitedSep 24, 2025
K251734 · ProVee Video Processing Unit (PV-003)
- Univlabs Technologies Private LimitedAug 11, 2025
K250462 · UL UHD-Clear View 4K Camera System (UL-3Chip 4K)
- Shenzhen Sophway Technology Co., Ltd.Jul 1, 2025
K250204 · Endoscopic Camera System
- Fujifilm CorporationApr 11, 2025
K243260 · FUJIFILM Processor EP-8000; FUJIFILM Endoscope Model EG-860R
- Olympus Corporations of the AmericasApr 2, 2025
K242067 · VISERA ELITE III VIDEO SYSTEM CENTER OLYMPUS OTV-S700; VISERA ELITE III LED LIGHT SOURCE OLYMPUS CLL-S700; 4K CAMERA HEAD OLYMPUS CH-S700-XZ-EA
- Guangzhou Red Pine Medical Instrument Co., Ltd.Feb 7, 2025
K243321 · Endoscopic Video Image Processor (RP-IPD-V1000F)