Tray, Biopsy, Without Biopsy Instruments
Gastroenterology, Urology · 21 CFR 876.4730
Recent MDR events
6 adverse event reports on file
- ENDOCHOICE COMPLIANCE ENDOKITFeb 15, 2018InjuryBoston Scientific/endochoiceMW5075302
- ENDOCHOICEFeb 15, 2018InjuryBoston Scientific/ EndochoiceMW5075303
- ADULT LUMBAR PUNCTURE TRAYFeb 9, 2017InjuryCarefusion/ BdMW5067829
- UCH CUSTOM BLOCK TRAYMay 9, 2013MalfunctionBusse Hosp DisposablesMW5030153
- BASIC BIOPSY TRAYFeb 10, 2000InjuryAllegiance Healthcare Corp.1423507-2000-00039
- JAMSHIDI BONE MARROW BIOPSY TRAYJan 19, 1993InjuryBaxter Healthcare3846
Event type breakdown
Recalls
14 recall events for this product code
- American Contract Systems Inc.Open, ClassifiedOct 16, 2025
Basic Biopsy Tray, Item Number NMBP44L
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Z-0170-2026 · View FDA record ↗ - MEDLINE INDUSTRIES, LP - NorthfieldOpen, ClassifiedSep 6, 2024
Medline Convenience kits labeled as:
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Z-3004-2024 · View FDA record ↗ - American Contract Systems, Inc.Open, ClassifiedAug 7, 2024
American Contract Systems Biopsy Drape Pack convenience kit
Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber, which may have been compromised in sterility
Z-2523-2024 · View FDA record ↗ - American Contract Systems, Inc.Open, ClassifiedJan 26, 2024
Custom procedural convenience kits and trays, gastroenterological & urological, labeled as:
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
Z-0821-2024 · View FDA record ↗ - MEDLINE INDUSTRIES, LP - NorthfieldOpen, ClassifiedAug 3, 2023
Custom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following:
Potential for the labeled sterile ultrasound gel component within packs to not meet sterility specifications.
Z-2310-2023 · View FDA record ↗ - MEDLINE INDUSTRIES, LP - NorthfieldOpen, ClassifiedJul 20, 2023
Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) FETOSCOPY CDS-LF, Model Number CDS840215O;
Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
Z-2173-2023 · View FDA record ↗ - MEDLINE INDUSTRIES, LP - NorthfieldOpen, ClassifiedFeb 23, 2023
Medline Sterile Oncology Port Access Tray, Item #DYNDC2239, 20 sterile kits per case, containing Ethicon Biopatch Protective Disk with CHG, Disk with Chlorhexid
The convenience trays had improper placement of some of the components that were in the nonsterile portion of the tray, thus, not receiving sterilization.
Z-1161-2023 · View FDA record ↗ - MEDLINE INDUSTRIES, LP - NorthfieldOpen, ClassifiedFeb 6, 2023
Sterile Procedural Trays, labeled as the following:
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
Z-1073-2023 · View FDA record ↗ - Custom Healthcare Systems, Inc.TerminatedFeb 7, 2013
Custom Breast Biopsy Tray, Catalog # B 9-16261, packaged in bags, 1 tray/bag, 20 bags/case.
Surgical procedure trays are packed with pharmaceuticals for injection that may contain particulate matter.
Z-0796-2013 · View FDA record ↗ - Custom Healthcare Systems, Inc.TerminatedFeb 7, 2013
Custom Breast Biopsy Tray w/Sodium Chloride, Catalog # B9-15263A, packaged in bags, 1 tray/bag, 20 bags/case.
Surgical procedure trays are packed with pharmaceuticals for injection that may contain particulate matter.
Z-0797-2013 · View FDA record ↗
Registered firms
27 firms currently registered for this product code
Medical Action Industries, Inc. 306
ARDEN, NC, US
Productos Medline S.a. De C.v.
NUEVO LAREDO Tamaulipas, MX
Lsl Healthcare
Niles, IL, US
Avid Medical, Inc. A Medical Action Ind. Company 305
TOANO, VA, US
Hk Surgical, Inc.
SAN CLEMENTE, CA, US
Tessy Plastics Shanghai Co., Ltd
SHANGHAI, CN
Medline Industries, Lp - Laredo
LAREDO, TX, US
Devicor Medical Products, Inc.
Cincinnati, OH, US
Sterigenics U.s., Llc
Ontario, CA, US
Diversatek Healthcare Inc
Milwaukee, WI, US
Therapak, Llc
Oakwood, GA, US
Nuevo Laredo Directo Al Cliente
Nuevo Laredo Tamaulipas, MX
Xinhua Surgical Instrument Co. Ltd.
ZIBO Shandong, CN
Sterigenics Us Llc
Grand Prairie, TX, US
Busse Hospital Disposables
HAUPPAUGE, NY, US
Medline Industries, Lp - Spt
Waukegan, IL, US
Midwest Sterilization Corp.
LAREDO, TX, US
Medline Soluciones Medicas
Nuevo Laredo Tamaulipas, MX
Medline Industries, Lp - Mexicali (mxc)
Mexicali Baja California, MX
Falcon Surgical Co. (pvt.) Ltd.
Sialkot Punjab, PK
United States Endoscopy Group, Inc.
Mentor, OH, US
Dynamic Diagnostics Incorporated
Plymouth, MI, US
Midwest Sterilization Corp.
JACKSON, MO, US
Honduras Kitting, S.a.
Villanueva Cortes, HN
510(k) clearances
2 clearances for this product code