Recalls for Pump, Portable, Aspiration (manual Or Powered)
21 on file
- R A Medical ServicesOpen, ClassifiedSep 20, 2023
Inadequate documentation of acceptance activity
- Ohio Medical CorporationTerminatedSep 7, 2018
The care-e-vac(R)3 Battery/AC Powered Portable Aspirator battery terminal may come in contact with the pump causing an electric short leading to damage to wiring and resulting in loss of function or an inoperable unit under battery power.
- Ohio Medical CorporationTerminatedDec 29, 2010
The D-14 diode on the circuit board may fail, resulting in a damaged, overheated circuit board. Also a screw used to attach one of the feet of the unit may damage the battery inside the device if the screw is over-tightened.
- Ohio Medical CorporationTerminatedJun 7, 2010
Ohio Medical has received a report of an explosive event with a MoblVac Portable Woundcare Pump housing where parts were separated while in an enclosed nylon transport bag during use.
- Ohio Medical CorporationTerminatedJun 7, 2010
Ohio Medical has received a report of an explosive event with a MoblVac Portable Woundcare Pump housing where parts were separated while in an enclosed nylon transport bag during use.
- Smith And Nephew, Inc. Wound Management DivisionTerminatedOct 27, 2008
External battery chargers used with the Version 29 VISTA Negative Pressure Wound Therapy pumps are failing to properly charge the pump's battery.
- Neomedix CorpTerminatedSep 29, 2008
Failure of irrigation flow: The pinch valve in a few pumps may not consistently open to allow irrigation flow. In over 450 surgeries, the firm confirmed one instance of a valve that did not open. The electrically activated pinch valve in the NeoMedix Trabectome I/A pump activates irrigation flow to the eye during Trabectome surgery.
- Medtronic NeuromodulationTerminatedSep 4, 2008
The Medela Vario 18 Vacuum Pump used with the Medtronic Bravo pH Monitoring System. The Vacuum Pump is used to provide the vacuum pressure during Capsule attachment. Some Medela Vario 18 Vacuum Pumps were provided with an incorrect suction Jar Lid that did not have the correct conical connection for the Vacuum Tube.
- Smith And Nephew, Inc. Wound Management DivisionTerminatedAug 6, 2008
Incorrect labeling. Kit labels may incorrectly state that a flat drain is contained as a kit component while some kits may contain a round drain.
- Haemonetics CorporationTerminatedOct 12, 2006
Faulty circuit board may short circuit and cause electric shock to operator
- S S C O R IncTerminatedJan 27, 2005
Reports from medical professionals that SSCOR suction device has not performed as intended. Firm's investigation disclosed that the condition in questions is caused by cracked regulators that result in inadequate suction.
- S S C O R IncTerminatedJan 27, 2005
Reports from medical professionals that SSCOR suction device has not performed as intended. Firm's investigation disclosed that the condition in questions is caused by cracked regulators that result in inadequate suction.
- S S C O R IncTerminatedJan 27, 2005
Reports from medical professionals that SSCOR suction device has not performed as intended. Firm's investigation disclosed that the condition in questions is caused by cracked regulators that result in inadequate suction.
- S S C O R IncTerminatedJan 27, 2005
Reports from medical professionals that SSCOR suction device has not performed as intended. Firm's investigation disclosed that the condition in questions is caused by cracked regulators that result in inadequate suction.
- S S C O R IncTerminatedJan 27, 2005
Reports from medical professionals that SSCOR suction device has not performed as intended. Firm's investigation disclosed that the condition in questions is caused by cracked regulators that result in inadequate suction.
- S S C O R IncTerminatedJan 27, 2005
Reports from medical professionals that SSCOR suction device has not performed as intended. Firm's investigation disclosed that the condition in questions is caused by cracked regulators that result in inadequate suction.
- Precision Medical, Inc.TerminatedDec 15, 2004
Battery charger can be overcharged and overheat
- Precision Medical, Inc.TerminatedDec 15, 2004
Battery charger can be overcharged and overheat
- Ohio Medical CorporationOpen, ClassifiedUnknown
Increase in premature device failures
- MicroAire Surgical Instruments, LLCTerminatedUnknown
Complaints that the cannula bent/broke during surgery.
- Medela AG Medical TechnologyTerminatedUnknown
Incorrect expiration date of "2023-07-24" on label of sterile tubing. Correct expiration date should be "2021-07-24".