RESOLVE HALO
Received Apr 15, 2020 · Event occurred Dec 30, 2019
Report 0001836248-2020-00002 · MDR key 9961409
Device
Product problems
- Malposition of Device
Patient
Not reported
- Seizures
- Swelling
- Brain Injury
- Seizures
- Swelling
- Brain Injury
Narrative
Additional Manufacturer Narrative
BASED ON INFORMATION KNOWN IT IS LIKELY NO FAILURE OR MALFUNCTION OCCURRED WITH THE DEVICE. THE EVENT LIKELY OCCURRED TO PIN PENETRATION, WHICH MAY HAVE BEEN DUE TO IMPROPER APPLICATION/MAINTENANCE OF THE PIN TIGHTNESS OR MAY HAVE BEEN DUE TO A FALL/IMPACT INCIDENT THAT WAS NOT RELATED TO A PRODUCT MALFUNCTION.
Description of Event or Problem
THE PATIENT HAD A HALO VEST FITTED, THIS WAS READJUSTED TO IMPROVE ALIGNMENT A FEW DAYS LATER. THE PATIENT WAS SUBSEQUENTLY READMITTED WITH A SEIZURE. IMAGING CONFIRMED THE POSTERIOR HALO PINS HAD BREACHED THE SKULL BILATERALLY AND THERE WAS MINOR BRAIN SWELLING. THE PATIENT WAS ADMITTED TO HOSPITAL, THE HALO VEST WAS REMOVED, AND THE PATIENT TREATED WITH ANTICONVULSANTS AND ANTIBIOTICS.
Remedial action
- Patient Monitoring