inforMED
MalfunctionHTE

PERIOSTEAL ELEVATOR 6MM CURVED BLADE-STRAIGHT EDGE

Received Mar 19, 2020 · Event occurred Jan 1, 2020

Report 2939274-2020-01364 · MDR key 9853886

Device

Generic name

Elevator

Model number

399.40

Catalog number

399.40

Product problems

  • Break
  • Break

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

INITIAL REPORTER IS SYNTHES SALES REPRESENTATIVE. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2020, THE SHAFT BROKE IN TWO PARTS. THE DISTAL PART WAS DISCARDED. NO FURTHER INFORMATION PROVIDED. THIS REPORT IS FOR ONE (1) PERIOSTEAL ELEVATOR 6MM CURVED BLADE-STRAIGHT EDGE. THIS IS REPORT 1 OF 1 FOR (B)(4).

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: H3, H6: VISUAL INSPECTION: THE PERIOSTEAL ELEVATOR 6MM CURVED BLADE-STRAIGHT EDGE WAS RECEIVED AT US CUSTOMER QUALITY (CQ). VISUAL INSPECTION OF THE COMPLAINT DEVICE SHOWED THE TIP WAS BROKEN OFF. THERE WAS NO OBVIOUS MATERIAL DEFICIENCY. DIMENSIONAL INSPECTION: A DIMENSIONAL INSPECTION WAS NOT PERFORMED DUE TO POST-MANUFACTURING DAMAGE. DOCUMENT/SPECIFICATION REVIEW: CURRENT REVISION WAS REVIEWED. NO DESIGN ISSUES OR DISCREPANCIES WERE IDENTIFIED. INVESTIGATION CONCLUSION: THIS COMPLAINT IS CONFIRMED AS THE TIP WAS BROKEN OFF. NO DEFINITIVE ROOT CAUSE COULD BE DETERMINED BASED ON THE PROVIDED INFORMATION. THERE WAS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE CONTRIBUTED TO THE COMPLAINT. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. H11 CORRECTED DATA: D10. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.