MOBILE CART CAMERA

Received Mar 9, 2020

Report 9612330-2020-00003 · MDR key 9805143

Device

Generic name

Mobile Cart Camera

Model number

10295

Product problems

  • Device Damaged Prior to Use
  • Defective Device
  • Device Damaged Prior to Use
  • Defective Device

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • Not Applicable
  • No Known Impact Or Consequence To Patient
  • Not Applicable

Narrative

Description of Event or Problem

MOBILE CART CAMERA FELL ON THE EEG TECHNICIAN HEAD.

Additional Manufacturer Narrative

UPDATE 07TH MAY 2020: INVESTIGATION RESULTS & FINDINGS (NATUS COMPLAINT REF. # 1-702493582). RISK REVIEW: PER DOC-010378 - EEG/PSG RISK ANALYSIS, HAZARD ID - 6.4 IDENTIFIES "SYSTEM CART OR PERIPHERALS FALL OR TIP ONTO PATIENT OR USER WHEN CART IS IDLE, IN MOTION, OR DURING ATTEMPT TO TRAVERSE A THRESHOLD OR OBSTACLE." THE RISK IS CONSIDERED MODERATE. UPON EVALUATION OF THE UNIT, IT WAS NOTED THAT THE MOUNTING TREAD ON BOTH SIDES OF THE CAMERA ARE COMPLETELY DESTROYED WHICH WOULD HAVE BEEN CAUSED BY MISALIGNING THE CAMERA AND POLE WHEN THE CAMERA WAS MOUNTED. THE DESTRUCTION OF THE BOTTOM TREAD SUGGESTS THAT AN ATTEMPT WAS MADE TO MOUNT IT ON THE POLE. THE LACK OF A SECURE, RIDGED ASSEMBLY BETWEEN THE CAMERA AND CARTS POLE COULD RESULT IN THE CAMERA FALLING OFF THE CART. A REPLACEMENT UNIT HAS BEEN SENT TO THE CUSTOMER. THE COMPLAINT AS REPORTED WAS VERIFIED. THIS HAS BEEN DEEMED AN ISOLATED INCIDENT AND WILL CONTINUE TO BE MONITORED FOR FUTURE OCCURRENCE.

Description of Event or Problem

MOBILE CART CAMERA FELL ON THE EEG TECHNICIAN HEAD.

Additional Manufacturer Narrative

UPDATE 09TH APRIL 2020 MANUFACTURERS INTERNAL REPORT REF. # (B)(4). INVESTIGATION RESULTS & FINDINGS THE SUSPECT DEVICE HAS BEEN RETURNED FOR EVALUATION. INVESTIGATION IS IN PROGRESS.

Additional Manufacturer Narrative

MOBILE CART CAMERA HAS FALLEN ON THE EEG TECHNICIAN HEAD.

Description of Event or Problem

MOBILE CART CAMERA FELL ON THE EEG TECHNICIAN HEAD.