COMBUR 5-TEST HC 10 STR.
Received Mar 4, 2020 · Event occurred Feb 11, 2020
Report 1823260-2020-00625 · MDR key 9785787
Device
Generic name
Urine Test Strips
Manufacturer
Roche DiagnosticsModel number
NACatalog number
11896954176
Lot number
42457401
Product problems
- False Positive Result
- False Positive Result
Patient
Not reported
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
THE REPORTER'S PRODUCT WAS REQUESTED FOR INVESTIGATION. ONE TEST STRIP BOX PACKAGE WAS RETURNED FOR INVESTIGATION. THE BOX DID NOT CONTAIN A TEST STRIP VIAL. THE PACKAGE SHOWED NO DAMAGES. THE RETENTION MATERIAL OF TEST STRIP LOT 424574 WAS VISUALLY CHECKED. THE RETENTION MATERIAL SHOWED NO DISCOLORATIONS. THE MATERIAL FULFILLS REQUIREMENTS.
Description of Event or Problem
THE INITIAL REPORTER STATED THEY RECEIVED A FALSE POSITIVE NITRITE RESULT FOR ONE PATIENT SAMPLE TESTED WITH COMBUR 5 TEST STRIPS. WITH THE FIRST TEST STRIP IN THE VIAL, THE REPORTER NOTICED A POSITIVE DISCOLORATION IN THE NITRITE FIELD OF THE TEST STRIP. THIS STRIP WAS DISPOSED OF IMMEDIATELY. NO PROBLEMS WERE OBSERVED WITH OTHER TEST STRIPS IN THE VIAL. THE REPORTER STATED THAT OTHER MEASUREMENTS PERFORMED WITH TEST STRIPS FROM THE VIAL ALWAYS SHOWED A POSITIVE NITRITE RESULT. ONE PATIENT SAMPLE RESULTED WITH A POSITIVE NITRITE VALUE WHEN TESTED USING THE TEST STRIPS. WHEN THIS PATIENT SAMPLE WAS REPEATED AT THE DOCTOR'S OFFICE AND PEDIATRICIAN'S OFFICE USING UNKNOWN TESTING METHODS, THE NITRITE RESULTS WERE NEGATIVE. THE NEGATIVE VALUES WERE BELIEVED TO BE CORRECT.
Additional Manufacturer Narrative
THE REPORTER'S TEST STRIPS WERE PROVIDED FOR INVESTIGATION. THE STRIPS WERE CHECKED VISUALLY AND WERE TESTED WITH NATIVE URINE AND A NITRITE DILUTION SERIES. THE RESULTS OF THE MEASUREMENTS FULFILLED REQUIREMENTS. NO FALSE POSITIVE RESULTS WERE OBSERVED.