5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 40MM
Received Mar 3, 2020 · Event occurred Feb 18, 2020
Report 3006460162-2020-00038 · MDR key 9779566
Device
Generic name
Response 5.5/6.0 System
Manufacturer
Orthopediatrics, CorpModel number
N/ACatalog number
00-1300-0640
Lot number
UNKNOWN
Product problems
- Fracture
- Fracture
Patient
16 YR
- Pain
- Pain
Narrative
Additional Manufacturer Narrative
REFERENCE: (B)(4). CONCOMITANT MEDICAL PRODUCTS: 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 40MM, ITEM NUMBER: 00-1300-0640, LOT NUMBER: UNKNOWN. COMPLAINT SAMPLE WAS NOT RETURNED FOR EVALUATION, SCRAPPED BY THE HOSPITAL. REPORTED EVENT WAS CONFIRMED BY REVIEW OF CUSTOMER SUBMITTED RADIOGRAPHS AND CT FILMS. DHR REVIEW WAS UNABLE TO BE PERFORMED AS THE LOT NUMBER OF THE INVOLVED DEVICE IS UNKNOWN. RISK MANAGEMENT FILE REVIEW WAS DEEMED APPROPRIATE. REVIEW OF THE COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION IS REQUIRED AS THERE WERE NO TRENDS IDENTIFIED. ROOT CAUSE WAS UNABLE TO BE DETERMINED.
Description of Event or Problem
IT WAS REPORTED THE PATIENT PRESENTED COMPLAINING OF PAIN AT THE LOWER END OF THE IMPLANTED SPINAL CONSTRUCT. RADIOGRAPHS AND CT SCAN SHOWED A LACK OF FUSION AT THE LOWER LEVEL. TWO PEDICLE SCREWS WERE FOUND TO BE BROKEN AND A REVISION SURGERY WAS REQUIRED.