QUICKVUE INFLUENZA A+B
Received Feb 27, 2020 · Event occurred Jan 22, 2020
Report 2024674-2020-00002 · MDR key 9766131
Device
Generic name
Quickvue Flu Test
Manufacturer
Quidel CorporationModel number
20183Catalog number
20183
Lot number
705526
Product problems
- False Negative Result
- False Negative Result
Patient
61 YR
- Death
- Sudden Cardiac Death
- Death
- Sudden Cardiac Death
Narrative
Additional Manufacturer Narrative
MANUFACTURER TESTED IN-HOUSE INFLUENZA A AND INFLUENZA B LOW-LEVEL POSITIVE STANDARD ON RETAINS AND RETURNS OF THE AFFECTED PRODUCT. ALL TESTS YIELDED ACCURATE AND VALID RESULTS AT THE 10 MINUTE RESULT READ TIME, PRODUCT PERFORMED PROPERLY. REVIEW OF THE MANUFACTURING RECORDS FOUND NO ISSUES THAT WOULD LEAD TO A FALSE NEGATIVE RESULT. BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY AND NO CORRECTIVE ACTION IS REQUIRED.
Description of Event or Problem
VERY SICK PATIENT CAME INTO THE CLINIC ON JAN 22 AND TESTED NEGATIVE FOR FLU WITH QUICKVUE TEST. HE WAS PRESCRIBED AZITHROMYCIN BECAUSE THEY BELIEVED HE HAD PNEUMONIA. ON JAN 23, A CHECK-IN WAS PERFORMED AND HE STATED HE WAS FEELING BETTER. ON JAN 24TH, 9:45 AM, HE CALLED IN REPORTING TO FEEL WORSE AND DEHYDRATED. HE DROVE TO THE HOSPITAL (B)(6) MEDICAL CENTER AND ARRIVED TO THE ER AT 2:56 PM. HE TESTED POSITIVE FOR FLU A ON MOLECULAR TEST (SOLANA) AND DIED LATER THAT DAY DUE TO CARDIAC ARREST.