Narrative

Description of Event or Problem

GAMBRO RECEIVED A COMPLAINT ON 10/15/07 FROM A FACILITY WHO REPORTED THAT A PHOENIX MACHINE WAS LEAKING DURING TREATMENT. A GAMBRO TECH INSPECTED THE MACHINE AND FOUND THAT THE TUBING WAS DISCONNECTED BETWEEN PFS AND EVDS1 VALVE AND AT EVUFO VALVE. HE RECONNECTED THE TUBING AT THE PFS/EVDS1 VALVE AND THE EVUFO VALVE. HE RAN THE MACHINE THROUGH T1 TEST, A SIMULATED PT RUN AND AN ADR BLEACH DISINFECT CYCLE.