inforMED
InjuryHPZ

BIO-EYE

Received Jan 2, 2008 · Event occurred Dec 21, 2007

Report 2027377-2007-00005 · MDR key 974749

Device

Generic name

Implant, Eye Sphere

Model number

I00020S

Catalog number

I00020S

Lot number

976115C

Product problems

  • Explanted
  • Implant extrusion

Patient

40 YR

  • Surgical procedure

Narrative

Description of Event or Problem

PHYSICIAN'S OFFICE REPORTED THAT THE DOCTOR IMPLANTED A 20MM STERILE IMPLANT. THE PT NOW HAS AN "EXPOSED" IMPLANT. PT RETURNED FOR SURGERY TO EXPLANT DEVICE IN 2007.

Additional Manufacturer Narrative

EXPOSURES ARE AN ANTICIPATED POSSIBLE COMPLICATION WITH THIS TYPE OF SURGERY. A REVIEW OF THE LITERATURE FINDS EXPOSURE RATES RANGING FROM 2.5% TO 21.6%. DISCUSSION OF THESE CASES WITH THE REPORTING PHYSICIAN SUGGESTS A CORRELATION WITH CHANGES IN THE CLINIC'S WRAPPING PRACTICES. IT IS UNKNOWN WHETHER THERE IS A CAUSAL RELATIONSHIP. WE ARE ANTICIPATING THE RETURN OF THE EXPLANTED DEVICES FOR FURTHER INVESTIGATION.