inforMED
InjuryFII

FRESENIUS DIALYSIS DELIVERY SYSTEM

Received Jan 2, 2008 · Event occurred Dec 5, 2007

Report 2937457-2008-00001 · MDR key 973522

Device

Generic name

Hemodialysis Machine

Model number

2008K

Catalog number

NA

Lot number

NA

Product problems

  • Increase in Pressure

Patient

57 YR

  • Low Blood Pressure/ Hypotension
  • Dizziness
  • Hospitalization required

Narrative

Description of Event or Problem

A DIALYSIS FACILITY REPORTED, THAT A PATIENT C/O DIZZINESS AND BECAME HYPOTENSIVE ONE HOUR INTO A HEMODIALYSIS TREATMENT WAS GIVEN ONE LITER OF SALINE. SHE CONTINUED TO BE HYPOTENSIVE AND UNSTABLE EVEN AFTER RECEIVING AN ADDITIONAL 800 ML. OF SALINE. SHE WAS TAKEN TO THE HOSPITAL AND WAS ADMITTED FOR UNSTABLE HYPOTENSION. THE TRANSMEMBRANE PRESSURE INCREASED FROM 40 TO 120 DURING TREATMENT. THERE WAS NO OTHER MACHINE PROBLEM NOTED. THE PATIENT'S WEIGHT WAS 2 KG. BELOW HER DRY WEIGHT AFTER RECEIVING 1.8 LITERS OF SALINE.

Additional Manufacturer Narrative

DEVICE INVESTIGATION PERFORMED BY A FRESENIUS EQUIPMENT SPECIALIST SHOWED A UF COUNTER ERROR OF 21 ML (MAXIMUM ALLOWABLE IS 10 ML). THE FUNCTIONAL BOARD WAS REPLACED FOR THIS PROBLEM. THE EXCESS UF WAS NOT DUPLICATED.