inforMED
MalfunctionFBO

INNER SHEATH WITH CERAMIC BEAK

Received Feb 20, 2020

Report 9610617-2020-00024 · MDR key 9734564

Device

Generic name

Inner Sheath With Ceramic Beak

Model number

27050CA

Catalog number

27050CA

Lot number

NQ05

Product problems

  • Material Separation
  • Material Separation

Patient

Not reported

  • Tissue Damage
  • Tissue Damage

Narrative

Additional Manufacturer Narrative

THE DEVICE WAS EVALUATED BY THE MANUFACTURER, KARL STORZ IN (B)(4). AS PER THE EVALUATION: UPON EVALUATION IT WAS CONFIRMED THAT THE CERAMIC TIP IS BROKEN OFF. THE SHEATH SHOWS RESIDUALS OF THE ADHESIVE WHERE THE CERAMIC IS BROKEN OFF. THE TIP SHOWS A CONTINUOUS CRACK AND CERAMIC CHIPPING NEAR THE EDGE OF THE TIP. THE BREAK LINE STARTS AT THE EDGE OF THE TIP AND CONTINUES IN CIRCULAR WAY IN THE GLUING AREA. THE EDGE OT THE TIP SHOWS DISCOLORATION. THE DEVICE WAS MANUFACTURED IN JANUARY 2019. THE ROOT CAUSE MOST LIKELY WAS MECHANICAL DAMAGE TO THE CERAMIC TIP, WHICH LED TO THE CRACK AND SUBSEQUENTLY TO THE BREAKAGE. NO INDICATION FOR A MATERIAL OR MANUFACTURING RELATED ISSUE WAS FOUND DURING THE INVESTIGATION.

Description of Event or Problem

AS PER A VIGILANCE REPORT FILED WITH THE (B)(6) BY OUR PARENT COMPANY IN (B)(6), THE BEAK OF THE RESECTOSCOPE SHEATH BROKE OFF INSIDE THE BLADDER OF THE PATIENT. THE SURGEON WAS ABLE TO RETRIEVE THE BEAK ENTIRELY, BUT THE PATIENT SUFFERED SOME MILD URETHRAL TRAUMA AS A RESULT.