MalfunctionIRT
THERMOPHORE
Received Feb 12, 2020
Report 1832415-2020-11071 · MDR key 9702515
Device
Product problems
- Break
- Thermal Decomposition of Device
- Material Twisted/Bent
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
CUSTOMER STATED "THE LIGHT COMES ON BUT THERE IS NO HEAT." CUSTOMER DID NOT CLAIM INJURY. PRODUCT WAS RETURNED. THE INVESTIGATOR OBSERVED A BURN MARK ON THE INSIDE OF THE UNIT BY THE BUTTERFLY. THE INVESTIGATOR DETERMINED TWISTING OF THE WIRES CONNECTING THE CORD TO THE PAD CAUSED A BREAK IN THE CORD, WHICH LEAD TO THE BURN MARK. THIS IS A MANUFACTURES DEFECT. AFTER FURTHER INVESTIGATION, NO ADDITIONAL PRODUCT WAS IMPACTED AND NO ADDITIONAL ACTIONS WERE REQUIRED.