NIM® 2.0 MAINFRAME - PULSE
Received Feb 12, 2020 · Event occurred Aug 5, 2019
Report 1045254-2020-00082 · MDR key 9699163
Device
Generic name
Electromyograph, Diagnostic
Manufacturer
Medtronic Xomed Inc.Model number
8252201Catalog number
8252201
Lot number
44082600
Product problems
- Device Sensing Problem
- Device Sensing Problem
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
VISUALLY, THERE WAS NO SIGNIFICANT DAMAGE TO THE DEVICES THAT WOULD INDICATE A CAUSE FOR THE COMPLAINT. THE INTERFACE WAS PLUGGED INTO THE MAINFRAME AND THE MAINFRAME POWER WAS TURNED ON. THE MAINFRAME BOOTED UP WITH NO ISSUES OR ERRORS. THE TOUCHSCREEN WAS RESPONSIVE AND ACCURATE. A PATIENT SIMULATOR AND STIMULATION PROBE WERE PLUGGED INTO THE PATIENT INTERFACE. ALL PLUGS FIT SECURELY IN THEIR RESPECTIVE CONNECTORS. CHANNEL 1 WAS GREEN WITH A RESISTANCE OF 5.8KOHM ON THE POSITIVE SIDE AND 5.2KOHM ON THE NEGATIVE SIDE; CHANNEL 2 WAS GREEN WITH A RESISTANCE OF 5.2KOHM ON THE POSITIVE SIDE AND 5.9KOHM ON THE NEGATIVE SIDE. USING 1.0 MA STIMULATING CURRENT AND 150UV THRESHOLD; WHEN STIMULATING, THE SYSTEM CONSISTENTLY RETURNED AUDIO AND VISUAL WAVEFORM AND EVENT TONE AND STIMULATING CURRENT BETWEEN 0.89MA AND 0.90MA. SWITCHING THE TYPE OF ELECTRODE FROM SUBDERMAL TO EMG TUBE SHOWED NO CHANGE TO THE RESPONSES. THERE WAS NO INTERMITTENT BEHAVIOR WHILE MANIPULATING ALL CONNECTORS / PLUGS AND THEIR STRAIN RELIEFS. THE RETURNED MUTING CABLE WAS PLUGGED INTO ALL 4 PORTS ON THE BACK OF THE MAINFRAME AND IT DID NOT AFFECT STIMULATION. STIMULATION AND RESPONSES WERE CHECKED ON ALL 4 PROCEDURES LOADED IN THE SYSTEM WITH NO ISSUES (PAROTID, THYROID, MASTOID, AND ACOUSTIC NEUROMA). THERE WAS NO MALFUNCTION FOUND AS IT RELATES TO THE COMPLAINT. WITH NO FAULT FOUND WITH THE RETURNED DEVICE, THE INFORMATION MOST LIKELY INDICATES AN ISSUE WITH THE SET UP OR THE DEVICE WAS USED IN CONFLICT WITH THE USER¿S GUIDE. CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID: 8252210, SERIAL/LOT #: (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem
A HEALTHCARE PROVIDER (HCP) REPORTED THAT THE NERVE MONITORING DEVICES MALFUNCTIONED. THERE WAS NO NEUROLOGICAL SIGNAL DURING EXPLORATION OF THE RECURRENT NERVE. THERE WAS NO KNOWN PATIENT IMPACT.