PHILIPS M1026B ANESTHETIC GAS MODULE

Received Feb 7, 2020

Report 3007409280-2020-00040 · MDR key 9684663

Device

Generic name

Anesthetic Gas Module

Model number

FGPH1026B

Product problems

  • Complete Loss of Power
  • Complete Loss of Power

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. DEVICE SENT IN DUE TO THE DEVICE SHUTTING DOWN IN THE MIDDLE OF A PROCEDURE. INCOMING VISUAL INSPECTION FOUND NO ANOMALIES. DEVICE EVALUATION FOUND NO PROBLEM. THE DEVICE WAS CALIBRATED, THE CASE AND FRONT/REAR CONNECTORS WERE INSPECTED. THE CIRCUIT BOARDS WERE INSPECTED. THE REPORTED EVENT WAS UNABLE TO BE DUPLICATED AND THERE WAS NO PROBLEM FOUND; THEREFORE, A ROOT CAUSE CANNOT BE DETERMINED. THIS TYPE OF EVENT WILL CONTINUE TO BE MONITORED. IT SHOULD BE NOTED THAT THIS IS BEING FILED BASED ON RETROSPECTIVE REVIEW.

Description of Event or Problem

REPORTEDLY, POST REPAIR, THE DEVICE FAILED DURING A PROCEDURE AS IT SHUT OFF. THERE WAS NO REPORT OF PATIENT HARM OR INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.