COMBUR-10-M 100 STR
Received Feb 4, 2020 · Event occurred Jan 13, 2020
Report 1823260-2020-00286 · MDR key 9664089
Device
Generic name
Urine Test Strips
Manufacturer
Roche DiagnosticsModel number
COMBUR-10-MCatalog number
11379208173
Lot number
41516505
Product problems
- Non Reproducible Results
- Non Reproducible Results
Patient
8 YR
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
THIS EVENT OCCURRED IN (B)(6).
Description of Event or Problem
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT NITRITE RESULTS FOR ONE PATIENT SAMPLE TESTED WITH COMBUR 10 M TEST STRIPS ON A COBAS U 411 ANALYZER. THE REPORTER SUSPECTS THE UROBERRY MEDICATION THE PATIENT WAS TAKING INTERFERED WITH THE TEST METHOD. THE SAMPLE INITIALLY RESULTED WITH A POSITIVE NITRITE VALUE ACCOMPANIED BY A DATA FLAG WHEN TESTED ON THE U 411 ANALYZER. THE SAMPLE WAS REPEATED ON THE U 411 ANALYZER, RESULTING WITH A NEGATIVE NITRITE VALUE. THE SAMPLE WAS ALSO TESTED MANUALLY USING A COMPETITOR URINE TEST STRIP (SPINREACT) AND THE NITRITE RESULT WAS POSITIVE. THE U 411 ANALYZER SERIAL NUMBER IS (B)(4).
Additional Manufacturer Narrative
THE RETENTION MATERIAL OF LOT 41516500 WAS MEASURED ON A RETENTION COBAS U411 AND ALSO CHECKED BY VISUAL READING WITH NATIVE URINE AND A NITRITE DILUTION SERIES. THE RETENTION MATERIAL SHOWED NO FALSE POSITIVE RESULTS AND FULFILLED REQUIREMENTS.