inforMED
MalfunctionCDX

ELECSYS T4 ASSAY

Received Jan 29, 2020

Report 1823260-2020-90035 · MDR key 9641140

Device

Generic name

Radioimmunoassay, Total Thyroxine

Manufacturer

Roche Diagnostics

Model number

T4

Catalog number

07027885190

Lot number

ASKU

Product problems

  • High Test Results
  • High Test Results

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

FOR THE PENDING EVENT, INVESTIGATION OF THE PATIENT SAMPLE FOUND AN INTERFERENCE TO STREPTAVIDIN AND AN IMMUNOGLOBULIN THAT REACTS WITH THE REAGENT WHICH AFFECTS THE SAMPLE RESULTS. THIS INTERFERENCE IS DOCUMENTED IN PRODUCT LABELING FOR THE ASSAY: IN RARE CASES, INTERFERENCE DUE TO EXTREMELY HIGH TITERS OF ANTIBODIES TO ANALYTE-SPECIFIC ANTIBODIES, STREPTAVIDIN OR RUTHENIUM CAN OCCUR. THESE EFFECTS ARE MINIMIZED BY SUITABLE TEST DESIGN.

Additional Manufacturer Narrative

THE INVESTIGATION OF THE EVENT IS CURRENTLY ONGOING. THE DEVICE WAS NOT RETURNED FOR INVESTIGATION. THIS DEVICE IS NOT LABELED FOR SINGLE USE AND IS NOT REPROCESSED OR REUSED.

Description of Event or Problem

THIS REPORT SUMMARIZES <NOE>1</NOE> MALFUNCTION EVENT. QUESTIONABLE HIGH ELECSYS T4 ASSAY RESULTS WERE GENERATED BY A COBAS 6000 E601 MODULE. THE EVENT INVOLVED ONE PATIENT SAMPLE. THE PATIENT'S AGE WAS REQUESTED, BUT NOT PROVIDED. THE PATIENT'S WEIGHT WAS REQUESTED, BUT NOT PROVIDED. THE PATIENT IS MALE. THE PATIENT'S RACE WAS REQUESTED, BUT NOT PROVIDED. THE PATIENT'S ETHNICITY WAS REQUESTED, BUT NOT PROVIDED.