Additional Manufacturer Narrative
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INVESTIGATION SUMMARY: LEVEL A INVESTIGATION - COMPLAINT EVALUATION / COMPLAINT HISTORY CHECK FOR THE EVENT(S) THAT OCCURRED. SEVERITY: S_1__; OCCURRENCE: A COMPLAINT HISTORY CHECK WAS PERFORMED AND THIS IS THE 2ND RELATED COMPLAINT FOR FOREIGN MATTER ON LOT # 9163840. INVESTIGATION SUMMARY: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH #9163840. ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. INVESTIGATION CONCLUSION: AS NO SAMPLES AND/OR PHOTO(S) WERE RECEIVED THE INVESTIGATION CONCLUDED: UNCONFIRMED: BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE AS NO SAMPLES OR PHOTOS WERE RETURNED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. ROOT CAUSE DESCRIPTION: ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED AS NO SAMPLES OR PHOTOS WERE RETURNED. RATIONALE: BASED ON THE INVESTIGATION, NO ADDITIONAL INVESTIGATION AND NO CAPA IS REQUIRED AT THIS TIME.
Description of Event or Problem
IT WAS REPORTED THAT SWAB ALCOHOL 100 BX 1200 (B)(6) HAD FOREIGN MATTER IN THE BLISTERS. THIS WAS DISCOVERED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE ALCOHOL PRESENTS THE SAME SPOTS AND HALOES THAT APPEAR IN THE BLISTERS.