inforMED
MalfunctionOLL

ENTACT SEPTAL STAPLER 3-PACK

Received Dec 26, 2019 · Event occurred Nov 29, 2019

Report 3006524618-2019-00617 · MDR key 9519553

Device

Generic name

Septal Stapler/absorbable Staples

Manufacturer

Arthrocare Corp.

Model number

601-00100

Catalog number

601-00100

Lot number

01908004

Product problems

  • Material Separation
  • Material Separation

Patient

Not reported

  • Foreign Body In Patient
  • Foreign Body In Patient

Narrative

Additional Manufacturer Narrative

H10: THE DEVICE REPORTED, USED IN TREATMENT, WAS NOT RETURNED FOR EVALUATION. A PICTURE OF THE DEVICE HAS BEEN SENT. PER THE CUSTOMER PHOTOS, THERE IS NO VISUAL IDENTIFICATION THAT THE ADHESIVE APPLICATION INTO THIS DEVICE WAS PERFORMED. THE RELATIONSHIP BETWEEN THE PRODUCT AND REPORTED WAS ESTABLISHED. A REVIEW OF MANUFACTURING RECORDS FOR THE REPORTED LOT NUMBER 01908004 FOUND NO NON-CONFORMANCES OR ANOMALIES DURING MANUFACTURING PROCESS RELATED TO THE REPORTED EVENT. A COMPLAINT HISTORY REVIEW FOUND RELATED FAILURES; THIS FAILURE MODE WILL BE TRENDED TO ASSESS FOR ANY NECESSARY CORRECTIVE ACTIONS. THE COMPLAINT WAS VERIFIED AND THE ROOT CAUSE FOR THIS EVENT WAS DETERMINED TO BE A MANUFACTURING PROCESS ERROR. MANUFACTURING RECORDS REVIEW WAS COMPLETED.

Description of Event or Problem

IT WAS REPORTED THAT, DURING A SEPTOPLASTY, THE ENTACT SEPTAL STAPLER FELL APART AFTER THE NEEDLE BROKE AND POPPED OUT. ALL FRAGMENTS WERE REMOVED USING SUCTION. THERE WAS A BACKUP DEVICE AVAILABLE. NO DELAY OR PATIENT INJURY WAS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.