NICOLET ELITE
Received Dec 23, 2019 · Event occurred Nov 27, 2019
Report 3010611950-2019-00077 · MDR key 9512785
Device
Generic name
Elite Auto Shut-off 200r Main
Manufacturer
Natus Neuro IncorporatedModel number
XMD008Catalog number
XMD008
Lot number
ECQ0529
Product problems
- Device Sensing Problem
- Environmental Compatibility Problem
- Device Sensing Problem
- Environmental Compatibility Problem
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
REFERENCE (B)(4). DEVICE RETURNED TO NATUS ON DECEMBER 26, 2019. PER DEPOT REPAIR NOTES: "REPLACED THE SPEAKER, THE FRONT HOUSING AND THE PROBE CABLE. INSTALLED A NEW BATTERY AND TESTED, THE UNIT PASSED ALL FUNCTIONAL TESTS. THE PROBE WAS EVALUATED AND THERE WERE NO PROBELMS FOUND WITH THE PROBE."
Additional Manufacturer Narrative
REFERENCE (B)(4). CUSTOMER STATES THERE IS EXCESSIVE NOISE WHEN TRYING TO HEAR THE FETAL HEARTBEAT. CUSTOMER WAS PROVIDED A QUOTE FOR REPAIR ON THE UNIT ON (B)(6), 2019. ADDITIONAL ATTEMPT TO COLLECT PRODUCT MADE ON (B)(6), 2019 WHERE CUSTOMER CONFIRMED THEY WERE STILL IN POSSESSION AND PLANNED ON RETURNING UNIT. ADVERSE EVENT QUESTIONNAIRE SENT TO CUSTOMER FOR ADDITIONAL FOLLOW UP ON PATIENT INVOLVEMENT ON (B)(6) AND (B)(6), 2019.
Description of Event or Problem
NOISE PREVENTS USER FROM BEING ABLE TO HEAR FETAL HEARTBEAT.