inforMED
MalfunctionGZI

CHATTANOOGA VECTRA GENISYS 2CH COMBO PKG US STD, 5CM APPL

Received Dec 23, 2019 · Event occurred Apr 11, 2019

Report 9616086-2019-00085 · MDR key 9510522

Device

Generic name

Stimulator, Neuromuscular, External Functional

Manufacturer

Djo, Llc

Model number

2761

Product problems

  • Intermittent Energy Output
  • Intermittent Energy Output

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

ONE VECTRA GEN 2C COMBO UNIT WAS RETURNED FOR EVALUATION. THE UNIT PASSED FUNCTIONAL AND SAFETY TESTING. ALL WAVEFORMS TESTED WITHIN TOLERANCE. CUSTOMER DID NOT SEND IN ANY LEADWIRES OR ELECTRODES FOR EVALUATION.

Description of Event or Problem

IT WAS REPORTED THAT THE "INTENSITY INTERMITTENTLY SURGES" DURING TREATMENT. THERE WAS NO REPORTED PATIENT HARM.