MalfunctionMIU

ALINITY S CHAGAS

Received Dec 9, 2019

Report 3002809144-2019-01044 · MDR key 9437641

Device

Generic name

Chagas

Manufacturer

Abbott Germany

Catalog number

06P08-45

Lot number

01166BE00

Product problems

  • High Test Results
  • High Test Results

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. AN EVALUATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. THIS REPORT IS BEING SUBMITTED ON AN INTERNATIONAL PRODUCT, ALINITY CHAGAS LN 06P08-45, WHICH HAS A SIMILAR PRODUCT IN THE US, LN 06P08-50.

Description of Event or Problem

THE CUSTOMER OBSERVED AN INCREASE IN FALSELY ELEVATED CHAGAS RESULTS ON THE ALINITY S ANALYZER. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6) INITIAL 6.364, REPEAT 2.428. SID (B)(6) INITIAL 1.358, REPEAT 1.362. SID (B)(6) INITIAL 0.928, REPEAT 1.162. THE SAMPLES WERE CONFIRMED NEGATIVE ON ANOTHER ANALYZER: SID (B)(6) INITIAL 0.013, REPEATS 0.013, 0.03. SID (B)(6) INITIAL 0.019, REPEATS 0.027, 0.06. SID (B)(6) INITIAL 0.028, REPEATS 0.029, 0.04. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Additional Manufacturer Narrative

AN INVESTIGATION WAS PERFORMED AND IDENTIFIED A DEFICIENCY FOR ALINITY S CHAGAS, LIST NUMBER 06P08-45, LOT NUMBER 01166BE00. AN INCREASE IN COMPLAINT ACTIVITY RESULTED IN IDENTIFYING A PRODUCT DEFICIENCY OF CERTAIN LOTS OF ALINITY S CHAGAS. HOWEVER, THE INVESTIGATION DETERMINED THAT THE IMPACTED LOTS WERE NOT DISTRIBUTED IN THE US, AND THEREFORE NO FURTHER ACTION IN THE US IS WARRANTED.