inforMED
MalfunctionCEK

TOTAL PROTEIN GEN.2

Received Dec 9, 2019 · Event occurred Nov 16, 2019

Report 1823260-2019-04320 · MDR key 9436771

Device

Generic name

Biuret (colorimetric), Total Protein

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

03183734190

Lot number

426862

Product problems

  • Non Reproducible Results
  • Non Reproducible Results

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THIS EVENT OCCURRED IN (B)(6).

Description of Event or Problem

THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH TP2 TOTAL PROTEIN GEN.2 ON A COBAS 6000 C (501) MODULE. NO INCORRECT RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE REPORTER SUSPECTS LIPEMIA MAY HAVE AFFECTED THE RESULT, BUT THE MEASURED LIPEMIA INDEX OF THE SAMPLE WAS BELOW THE INDEX LEVEL WHICH IS DOCUMENTED IN PRODUCT LABELING AS INTERFERING WITH THE ASSAY. THE SAMPLE INITIALLY RESULTED WITH A TOTAL PROTEIN VALUE OF 11.1 G/L. THE SAMPLE WAS REPEATED, RESULTING WITH A VALUE OF 26.5 G/L. THE SAMPLE WAS REPEATED AFTER HIGH SPEED CENTRIFUGATION, RESULTING WITH A VALUE OF 60 G/L. THE C 501 ANALYZER SERIAL NUMBER IS (B)(4).

Additional Manufacturer Narrative

THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.